LTL3000 User Manual / June 2020
18
DELTA
•
Place the LTL3000 all the way back on the Qd calibration target facing in the direction indicated.
•
P
ress ‘Calibrate Qd’ on the screen.
•
Wait for success confirmation.
To calibrate the Qd value, place the LTL3000 on top of the Qd calibration target
.
Qd Dark calibration (never):
The instrument is factory calibrated for the Qd-value and thus Qd should
not
be re-calibrated.
Qd calibration
requires
a Qd calibration marking (optional). See above
The instrument will warn about this fact before calibration
Note:
Be aware that it is
NOT
possible to recover the factory calibration if the Qd value is recalibrated.
Normally, it is sufficient to make a reference calibration of the LTL3000 instrument prior to conducting
measurements. However, if the instrument in rare occasions measures incorrectly it is recommended to do a
dark calibration.
To do a dark calibration follow these steps:
•
Select ‘Calibrate’ from the main menu.
•
Press ‘Qd Dark calibration’ on the screen.
•
Tilt the instrument backwards and make sure it is not pointing at anything within a few meters.
•
Press OK and wait for success confirmation.
•
Important:
Perform Reference Qd calibration (see above)