Create GP2 Program
⚫
25
3: Set Recording Intervals
On selecting a measurement, DeltaLINK automatically sets a
default recording interval, of one hour, as shown below.
Illustration: DeltaLINK, GP2 Program tab with a default Recording
interval set to 1 hour for each of the defined Measurements.
Chose a recording interval appropriate for your soil type.
For instance, if sandy it may drain quickly, in which case the
fastest recording rate possible is about once every 5 seconds for
one WET150 on a GP2. (62 sensors may take 50-55 seconds in
the standard, sequential mode). In a heavy clay you may wish
to start recording once per hour and wait and see how quickly
the soil responds to wetting and drying events.
4: Next Steps
Click on
Apply
to install the program in the logger.
Select the
Sensors
tab and click on
Read now
to see the
WET150 readings.
(Please note the GP2 Simulator does not work with SDI-12
sensors)
Select the
Logger
window and
Start
to start logging…
Select the
Dataset
window from which you can
Retrieve
and
display all stored readings.
See also the SDI-12 for GP2 User Manual.