How to Use the AP4
Page 12
CALIBRATION
PREPARATION
You need a Calibration Kit with:-
•
Calibration Plate in plastic envelope
•
Pack of filter pads
•
Tape
You also need:-
•
Scissors
•
Distilled water - 250 ml should be more than adequate (only a small amount is
needed - to wet a filter pad takes 5 ml )
•
Tissue paper or blotting paper
You may wish to use a barometer, such as the wristwatch-style one which we can
supply as an accessory. Without it a 100 hPa (mbar) pressure change, or a 1000 m
change in altitude will cause a 10 % error in measured resistance.
If the pressure does not change significantly between calibration and measurement,
then you can get away without a barometer.
See Pressure on page 31.