LPPYRHE16 

-   15   -

 

V2.3

 

6.2

 

R

EADING THE MEASURES WITH THE 

M

ODBUS

-RTU

 PROTOCOL

 

In MODBUS mode, you can read the values measured by the pyrheliometer through 

the function code 04h (Read Input Registers). The following table lists the quantities 

available with the appropriate register number and address: 

Number 

Address 

Quantity 

Format 

Internal temperature in °C (x10) 

16-bit Integer 

Internal temperature in °F (x10) 

16-bit Integer 

Solar radiation in W/m

2

 

16-bit Integer 

Status register: 

bit0=1 

 solar radiation measurement error 

bit2=1 

 configuration data error 

bit3=1 

 program memory error 

16-bit Integer 

Average values of the last 4 measurements 

16-bit Integer 

Signal generated by the sensor in µV/10 

[e.g.: 816 means 8160 µV, the resolution is 10 µV]

 

16-bit Integer 

Note: Register address = Register number - 1, as defined in the Modbus standard

.

 

O

PERATING MODE

: the pyrheliometer enters RS485 MODBUS-RTU mode after 10 sec-

onds from power on. In the first 10 seconds from power on the pyrheliometer does 

not reply to requests from the MODBUS master unit. After 10 seconds, it is possible to 

send MODBUS requests to the pyrheliometer. 

Содержание LPPYRHE16

Страница 1: ...Operating manual Pyranometer LPPYRHE16 www deltaohm com English Keep for future reference...

Страница 2: ...4 4 LPPYRHE16S CONNECTIONS 10 4 5 LPPYRHE16ACS CONNECTIONS 11 5 MEASUREMENT IN THE MODELS WITH ANALOG OUTPUT 12 5 1 LPPYRHE16 12 5 2 LPPYRHE16AC S 12 5 3 LPPYRHE16AV 12 6 RS485 MODBUS RTU OUTPUT 13 6...

Страница 3: ...vation seventh edition 2008 The pyrheliometer has a field of view of 5 in accordance with ISO 9060 2018 stand ard and WMO guide The pyrheliometer is available in the following versions LPPYRHE16 PASSI...

Страница 4: ...ld junction Thanks to the Seebeck effect the difference in temperature between hot and cold junction is converted into a Dif ference of Potential In order to reduce the variations of sensitivity depen...

Страница 5: ...order to reduce light scattering contribution Fig 2 3 scheme of principle LPPYRHE16 version with mV output Front opening Internal opening Sensor Field of view 2 arctan R d Slope angle arctan R r d The...

Страница 6: ...g a coin 2 Remove the cartridge perforated cap 3 Open the sachet containing silica gel supplied with the pyrheliometer 4 Fill the cartridge with the silica gel crystals 5 Close the cartridge with its...

Страница 7: ...es or obstructions intercept the sun s path during the day all year long To point the pyrheliometer the two holes in the front and back flange are used To properly align the instrument just make sure...

Страница 8: ...red versions internally there are surge protection devices connected to the housing Grounding the housing allows the correct protection functionality in particular against lightning 4 1 LPPYRHE16 CONN...

Страница 9: ...on the ordered output and requires external power supply 10 30 Vdc for 0 1 V and 0 5 V outputs 15 30 Vdc for 0 10 V output It is to be connected to a power supply and an instrument with voltage input...

Страница 10: ...Power supply positive Vdc Red 3 Not connected 4 RS485 A Brown 5 RS485 B White 6 Housing Cable shield SH Black 7 Not connected 8 Not connected Fig 4 4 LPPYRHE16S connections Before connecting the pyrhe...

Страница 11: ...The RS485 output is not isolated Connector Function Color 1 Power supply negative GND Blue 2 Power supply positive Vdc Red 3 Digital and analog ground SGND Black 4 RS485 A Brown 5 RS485 B White 6 Hou...

Страница 12: ...itivity of the pyrheliometer expressed in V Wm 2 5 2 LPPYRHE16AC S The 4 20 mA output signal corresponds to the 0 2000 W m2 irradiance range The irradiance Ee is obtained by measuring with a multimete...

Страница 13: ...to install the related USB drivers in the PC NOTES ON THE INSTALLATION OF UNSIGNED USB DRIVER before installing unsigned USB driver into operating systems starting from Windows 7 it is necessary to re...

Страница 14: ...data bits no parity 2 stop bits n 2 8 E 1 8 data bits even parity 1 stop bit n 3 8 E 2 8 data bits even parity 2 stop bits n 4 8 O 1 8 data bits odd parity 1 stop bit n 5 8 O 2 8 data bits odd parity...

Страница 15: ...eger 4 3 Status register bit0 1 solar radiation measurement error bit2 1 configuration data error bit3 1 program memory error 16 bit Integer 5 4 Average values of the last 4 measurements 16 bit Intege...

Страница 16: ...0 10 V LPPYRHE16AV Digital RS485 Modbus RTU LPPYRHE16 AC S Power supply 10 30 Vdc 4 20 mA 0 1 V and 0 5 V outputs 15 30 Vdc 0 10 V output 5 30 Vdc RS485 Modbus RTU output Connection 4 or 8 pole M12 c...

Страница 17: ...s depending on the envi ronment where the pyrheliometer works To exploit all the pyrheliometer features it is highly recommended that the calibration be checked annually 9 SAFETY INSTRUCTIONS General...

Страница 18: ...m CPM12 8DA 2 5 m CPM12 8DA 5 or 10 m CPM12 8DA 10 For LPPYRHE16ACS CP24 PC connecting cable for the RS485 MODBUS parameters configu ration of the LPPYRHE16 AC S pyrheliometers With built in RS485 US...

Страница 19: ......

Страница 20: ...warranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECH...

Отзывы: