HD50
- 8 -
V1.0
4
SOFTWARE
The parameters of the instrument (logging parameters, alarm thresholds, quantities
to be acquired, network settings, etc.) can be configured by connecting the instru-
ment to the PC via USB or Ethernet local network and using the
HD35AP-S
applica-
tion software (see the instructions of the software).
Some basic parameters (alarms, logging interval, user code, …) can also be set with
the
HDServer1
application software.
To download the data in a database, the
HD35AP-S
(it allows connecting one data log-
ger at a time) or
HDServer1
(it allows connecting several data loggers simultaneously)
application software and the
MySQL
database management system (included in the
application softwares packages) must be installed.
HD35AP-CFR21
SOFTWARE OPTION
The
HD35AP-CFR21
option allows, in addition to the features of the basic software
(both HD35AP-S and HDServer1), the protection of recorded data and instrument con-
figuration in response to
FDA 21 CFR part 11
recommendations. In particular be-
come available:
•
The traceability of activities (Audit Trail) performed with the software; for exam-
ple, which users connected and what changes were possibly made to the configu-
ration of the instrument.
•
The management of users access for the instrument configuration and viewing of
data in the database. Each user can be assigned a different password for using the
software. There are also three levels of access (Administrator, Super-user and stan-
dard User); for each level, the allowed operations can be defined.
The HD35AP-CFR21 option works with USB hardware key to be connected to any PC con-
nected to the same local network of the PC in which the basic software is installed.