31
Selectable print interval
Immediate or 1s, 5s, 10s, 15s, 30s, 1min,
2min, 5min, 10min, 15min, 20min, 30min
and 1hour
USB interface
Type
1.1 - 2.0 electrically isolated
Connections
Temperature probe input
with SICRAM module or TP47 module
8-pole male DIN45326 connector
Input for Oxygen probes
8-pole male DIN45326 connector
Serial interface and USB
8-pole MiniDin connector
Mains adapter (cod. SWD10)
2-pole connector (positive at centre)
12Vdc/1A
Measurement of the concentration of dissolved Oxygen
Measurement range
0.00…90.00mg/l
Resolution 0.01mg/l
Accuracy
±
0.03mg/l
±
1digit (60...110%, 1013mbar,
20...25°C)
Measurement of the concentration of dissolved Oxygen
Measurement range
0.0…600.0%
Resolution 0.1%
Accuracy
±
0.3%
±
1digit (in the range 0.0…199.9%)
±
1%
±
1digit (in the range 200.0…600.0%)
Automatic temperature compensation
0…50°C
Measurement of barometric pressure
Measurement range
0.0…1100.0mbar
Resolution 0.1mbar
Accuracy
±
2mbar
±
1digit between 18 and 25°C
±
(2mbar+0.1mbar/°C) in remaining range
Setting the salinity
Measurement range
0.0…70.0g/l
Resolution 0.1g/l
Temperature measurement with the sensor inside the dissolved Oxygen probe
Measurement range
0.0…+45.0°C
Resolution 0.1°C
Accuracy
±
0.1°C
Drift after 1 year
0.1°C/year
Temperature measurement by Instrument with Pt100 probe
Pt100 measurement range
-200…+650°C
Resolution 0.1°C
Accuracy
±
0.1°C
Drift after 1 year
0.1°C/year