Delta OHM HD3406.2 Скачать руководство пользователя страница 16

-      - 

16

probe is not present, at the manually-set temperature. In the lower line, the closest temperature 
compensated standard buffer value.

 

6.

 

Press the CAL key. The unit of measurement (

μ

S/cm or mS/cm) appears on the comment line.  

The central line shows the solution conductivity value at the measured temperature, or if the 
probe is not present, at the manually-set temperature. In the lower line, the closest temperature 
compensated standard buffer value.

  

If the measurement is in TDS, resistivity or salinity, by pressing CAL, the instrument goes 
automatically into conductivity calibration mode. 

7.

 

Press ENTER to confirm the displayed value. The cell constant nominal value (KCELL) and the 
set temperature coefficient 

α

T

 are displayed. Pressed repeatedly the ENTER key allows the cali-

bration on the point to be repeated, for example, in order to obtain a more stable value. 

8.

 

To end the probe calibration, press ESC. 

9.

 

Rinse the probe with water. If you are then going to perform low conductivity measurements, we 
recommend rinsing the probe using distilled or bidistilled water. 

 
The instrument is calibrated and ready for use. 
 
 

Manual calibration of conductivity using non memorized standard solutions 

 Manual calibration is possible at any calibration solution and temperature if it is within the instru-
ment measurement limits and provided that you know the solution's conductivity at the temperature 
at which the calibration is performed. Proceed as follows: 
1.

 

Turn the instrument on with the 

ON/OFF

 key. 

2.

 

Set the probe's cell constant by selecting it from the admitted values: 0.01, 0.1, 0.7, 1.0 or 10.0  

3.

 

Dip the conductivity meter cell into a known conductivity solution until the electrodes are cov-
ered with liquid. 

4.

 

Stir the probe lightly to remove any possible air inside the measurement cell. 

5.

 

Press MENU, and then FUNC/ENTER until the item ALPH appears. The temperature coeffi-
cient 

α

T

 is displayed. Note down the value displayed as it must be set again at the end of the 

procedure. Set the value to 

0.00

. This excludes the temperature compensation during the con-

ductivity measurement. 

6.

 

Measure the temperature by pressing °C/°F. According to the temperature detected, determine 
the calibration solution conductivity using the table specifying the conductivity according to 
temperature. 

7.

 

Select the conductivity measurement by pressing 

χ

-

-TDS. 

8.

 

Press the CAL key. The CAL  symbol is turned on. The unit of measurement (

μ

S/cm or 

mS/cm) appears on the comment line. If the calibration solution conductivity is sufficiently 
close (-30% to +50%) to one of the standard solutions, the secondary line displays the value.  
Otherwise it displays the calculated value according to current settings. In the central line the 
solution conductivity value is indicated according to the cell constant current settings.

 

9.

 

Use the arrows to select the conductivity value

 

determined at point 4 and confirm using EN-

TER. If the ERR indication appears, see the note below. 

Содержание HD3406.2

Страница 1: ...mation written in this manual and the instru ment that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this man...

Страница 2: ...2 Conductivity meter Thermometer HD3406 2...

Страница 3: ...o Fahrenheit 9 Function indicators 10 key in the menu confirms the current selection 11 Line for symbols and comments 12 Main display line 13 ON OFF AUTO OFF key turns the instrument on and off When p...

Страница 4: ...one or more of the 147 S 1413 S 12880 S or 111800 S cm conductivity standard solutions The displayed data can be stored datalogger and can be transferred to PC or serial printer thanks to the multi st...

Страница 5: ...ment ready for normal measure ment MIN cm min s 2 3 m m k During turning on should no probes be connected the last manually set temperature ap pears in the secondary line The unit of measurement symbo...

Страница 6: ...rks on command the sending of data to the serial port is performed each time the key is pressed Recording LOG is performed with one second intervals even if the interval is set to 0 With an interval f...

Страница 7: ...you can see an example of the measurement process with the Auto Hold function enabled In the following figure you can see an example of the measurement process with the Auto Hold function enabled A pr...

Страница 8: ...r to conduc tivity the indicated temperature is that measured at the maximum and minimum conductiv ity and is not the maximum and minimum temperature If the RCD MODE parameter is set to tp temperature...

Страница 9: ...onductivity liquid resistivity total dissolved solids salinity and temperature meas urements updating them with the acquisition of new samples The acquisition frequency is once a second Use the TDS ke...

Страница 10: ...e next the function is controlled by the Sleep_Mode_LOG parameter When the logging interval is less than one minute the logging instrument remains on with an interval of at least one minute it turns o...

Страница 11: ...is the same for both the instrument and the PC the communication will only work in this way Erase memory When pressed after the MENU key the SERIAL key permanently erases all the data contained in th...

Страница 12: ...12 Down Arrow When used in the menu it decreases the current variable value During measurement if the tem perature probe is not present it decreases the temperature value for conductivity compensation...

Страница 13: ...on of the temperature sensors is not required The probes are detected during turn on and this cannot be performed when the instrument is already on therefore if a probe is connected and the instrument...

Страница 14: ...ones for high values The indicative measurement range is reported in the following diagram Automatic or manual conductivity compensation The conductivity measurement refers to a standard temperature c...

Страница 15: ...ution 147 S cm 25 C 0 01 Molar KCl solution 1413 S cm 25 C 0 1 Molar KCl solution 12880 S cm 25 C 1 Molar KCl solution 111800 S cm 25 C Using one of these solutions the calibration is automatic the pr...

Страница 16: ...vity at the temperature at which the calibration is performed Proceed as follows 1 Turn the instrument on with the ON OFF key 2 Set the probe s cell constant by selecting it from the admitted values 0...

Страница 17: ...a long beep The instrument remains in calibration mode and maintains the previous calibration values The most frequent causes of error are due to the probe malfunctioning deposits dirt or to the stand...

Страница 18: ...at least 60mm the sensor is housed in the end part of the probe When measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip pr...

Страница 19: ...ide The wires coming from the conductivity cell are welded directly onto the 1 2 4 5 connector s pins On the left there are the 1 4 points on which the Pt100 or Pt1000 sensor wires must be welded The...

Страница 20: ...ns 1 and 4 Jumper between pins 2 and 5 4 electrode conductivity probe Internal view of the connector None As an alternative the Pt100 and Pt1000 sensors can be welded directly on the connector s pins...

Страница 21: ...Jumper between pins 1 and 4 Jumper between pins 2 and 5 4 electrode conductivity probe Pt100 tempera ture sensor Pt100 Internal view of the connector None Ensure the welds are clean and perfect Once...

Страница 22: ...ent is always very hard to perform It has high levels of uncertainty and depends on the ability of the operator 10 Temperature measurements on non metal surfaces usually require a great deal of time d...

Страница 23: ...ou work with a black Platinum cell new platinum coating of the electrode could be necessary Cell constant value not accepted Check that the standard solutions are in good condition that the probe s co...

Страница 24: ...full the instrument cannot store further data the memory space is full CAL blinking No calibration has yet been effected or the cell constant value has been modified in the menu please see the descri...

Страница 25: ...OG logging in progress LOG MEM FULL memory full LOG_CLRD memory data cleared LOG_DUMP_OR_ERAS transfer or erase data LOG_STOP logging complete LOG_STRT logging started MIN USE_UNIT_TO_ZERO SEC minutes...

Страница 26: ...ent issues a long beep and turns off In this case replace the batteries or use the external power supply in order to turn the instrument back on In order to avoid data loss the logging session is ende...

Страница 27: ...to direct sunlight The instrument may be exposed to a source of high temperature The instrument may be exposed to strong vibrations The instrument may be exposed to steam salt or any corrosive gas MA...

Страница 28: ...r where XY is the command code and cr is the Carriage Return ASCII 0D Command Response Description P0 Ping locks the instrument keyboard for 70 seconds P1 Unlocks the instrument keyboard S0 AT 21 3 10...

Страница 29: ...seconds WC0 Setting SELF off WC1 Setting SELF on Command characters are exclusively upper case characters Once a correct command is entered the instrument responds with when any wrong combination of c...

Страница 30: ...the LOG key the instrument logs the first data and turns off 15 seconds before the next logging instant it turns on again to capture the new sample and then turns off The data stored in the memory ca...

Страница 31: ...t by pressing SERIAL again See the description of the menu items on page 6 Connect the S print BT printer using cable HD2110CSP NOTES The print out is formatted across 24 columns Some keys are disable...

Страница 32: ...NU ENTER until the Baud Rate parameter select 38400 using the arrows confirm with ENTER The parameter remains in the memory until replacement of the batteries 4 Launch the DeltaLog9 application and pr...

Страница 33: ...tion could take a few minutes 17 The installation procedure is now complete the device will be detected on each new connec tion automatically In order to check if the entire operation was successful i...

Страница 34: ...egular functioning and operating security can be ensured only within the climatic conditions specified in the chapter TECHNICAL CHARACTERISTICS Do not use or store the instrument in ways and or places...

Страница 35: ...sorbed with instrument off 20 A Mains cod SWD10 Output mains adapter 100 240Vac 12Vdc 1A Security of memorized data Unlimited Time Date and time Schedule in real time Accuracy 1min month max departure...

Страница 36: ...Instrument Measurement range Kcell 0 01 Resolution Up to 1G cm Measurement range Kcell 0 1 Resolution Up to 100M cm Measurement range Kcell 1 Resolution 5 0 199 9 cm 0 1 cm 200 999 cm 1 cm 1 00k 19 9...

Страница 37: ...0 200 C Pt1000 measurement range 50 200 C Resolution 0 1 C Accuracy 0 25 C Drift after 1 year 0 1 C year Automatic manual temperature compensation 0 100 C con T 0 00 4 00 C Reference temperature 20 C...

Страница 38: ...K 0 01 0 04 S cm 20 S cm 0 120 C 2 electrode cell in AISI 316 Ultra pure water Measurement in closed cell SPT01G K 0 1 0 1 S cm 500 S cm 0 80 C 2 electrode cell in Platinum wire Probe material glass...

Страница 39: ...4 wires 50 200 C Class A TP47 1000 Pt1000 2 wires 50 200 C Class A TP87 100 Pt100 4 wires 50 200 C Class A TP87 1000 Pt1000 2 wires 50 200 C Class A Temperature drift 20 C 0 005 C TP47 Module for the...

Страница 40: ...nection of Pt100 4 wire and Pt1000 2 wire probes to instru ment series HD34 without amplifying electronics and linearization CONDUCTIVITY PROBES AND COMBINED CONDUCTIVITY AND TEMPERATURE PROBES SP06T...

Страница 41: ...cable with connector length 2 metres TP87 100 Pt100 sensor immersion probe Probe s stem 3mm length 70mm Connection cable 4 wires with connector length 1 metre TP87 1000 Pt1000 sensor immersion probe P...

Страница 42: ...mbined probes 4 wire Pt100 probes and 2 wire Pt1000 probes 18 WARNINGS AND OPERATING INSTRUCTIONS 22 Notes on conductivity measurement 22 INSTRUMENT SIGNALS AND FAULTS 24 LOW BATTERY WARNING AND BATTE...

Страница 43: ...43...

Страница 44: ...il pr sum d fectueux Dieser Garantieschein muss der Spedition beigelegt werden wenn das Ger t an das Kundendienstzent rum gesandt wird WICHTIG Die Garantie ist nur g ltig wenn dieser Abschnitt bis ins...

Отзывы: