23
10.2.2
R
ELATIVE HUMIDITY AND TEMPERATURE PROBES USING SICRAM MODULE
Model
Temperature
sensor
Application range
Accuracy
%RH
Temperature
%RH
Temp.
HP472ACR
Pt100 0...100%RH
-20°C…+80°C
±
1.5%RH (0...90%RH),
±
2.0%RH (elsewhere)
for T=15…35°C
---------
±
(1.5+1.5% reading)% RH
elsewhere in
temperature range
±0.3°C
HP572ACR
Thermocouple K 0...100%RH
-20°C…+80°C
±0.5°C
HP473ACR
Pt100
0...100%RH
-20°C…+80°C ±0.3°C
HP474ACR
Pt100
0...100%RH -40°C…+150°C ±0.3°C
HP475ACR
Pt100 0...100%RH -40°C…+150°C
±0.3°C
HP475AC1R
Pt100 0...100%RH -40°C…+180°C
±0.3°C
HP477DCR
Pt100 0...100%RH -40°C…+150°C
±0.3°C
HP478ACR
Pt100 0...100%RH -40°C…+150°C
±0.3°C
Common characteristics
Relative humidity
Sensor Capacitive
Temperature drift @20°C
max 0.02%RH/°C
Response time %RH at constant temperature
10sec (10
Æ
80%RH; air speed=2m/s)
Temperature with sensor Pt100
Temperature drift @20°C
0.003%/°C
Temperature with thermocouple K - HP572ACR
Temperature drift @20°C
0.02%/°C
For
HP 480
probe for measuring humidity in pipes, apply the above specifications with the
following exceptions:
HP 480
Temperature
Measuring range
-40 ... +60°C
Humidity
Dew Point TD
-40 ... +60°C DP
Environmental
conditions
Operating temperature
-40…+60°C
Operating pressure
16 bar max
10.2.3
4
WIRE
Pt100
AND
2
WIRE
Pt1000
P
ROBES
Model
Type
Application range
Accuracy
TP47.100.O
4 wire Pt100
-50…+250°C
1/3 Din
TP47.1000.O
2 wire Pt1000
-50…+250°C
1/3 Din
Common characteristics
Temperature drift @20°C
Pt100
0.003%/°C
Pt1000
0.005%/°C
Содержание HD2301.0R