The following table lists all error indications and information displayed by the instrument and
supplied to the user in different operating situations:
Display indication
Explanation
- - -
This appears in the display central line when a
temperature only probe is connected. In the lower line
the temperature is shown correctly.
PRBE_TYPE
Type of probe connected
BATT TOO LOW - CHNG NOW
Battery discharged - replace it immediately
CAL
LOST
Program error: it appears after turning on for a few
seconds. Contact the instrument's supplier.
CAL__11_ UP DOWN
Calibration at 11%RH, use the arrows to correct the
value proposed
CAL__33_ UP DOWN
Calibration at 33%RH, use the arrows to correct the
value proposed
CAL__75_ UP DOWN
Calibration at 75%RH, use the arrows to correct the
value proposed
ERR
This appears if the probe has already been detected by
the instrument, but is disconnected. At the same time an
intermittent beep is issued.
FUNC CLRD
max, min and average values cleared
NEW PROBE DET
This message appears when a probe is inserted into a
functioning instrument. Turn the instrument off and then
back on again.
NO_PRBE_ SER_NUM
The connected probe's serial number is absent
nonE
No selection
OVER
Measurement overflow: indicates that the probe is
measuring a value exceeding the measuring range.
PLS_EXIT >>> FUNC RES_FOR_
FACT ONLY
Please exit using ESC >>> function reserved to factory
calibration
PRBE_SER #### ####
Serial number #### #### of the connected probe
PROB
ERR
A probe with SICRAM module has been inserted when
not admissible for that specific instrument.
PROB COMM LOST
This appears if the probe has already been detected by
the instrument, but is disconnected. At the same time an
intermittent beep is issued.
PROB_CAL_EXIT_OR_75_OR_33_
OR_11
Probe calibration - exit or select 75, 33 or 11%RH
SYS
ERR
#
Instrument management program error. Contact the
instrument's supplier and communicate the numeric code
# reported by the display.