background image

-           - 

32

Calibration of the dissolved Oxygen probe 

The probe must be calibrated periodically using the DO9709/20 (for polarographic probe) or 
DO9709/21 (for galvanic probe) calibrator. 
The instrument checks the dissolved Oxygen probe efficiency. The "OFS_ERROR" message 
indicates that the probe is exhausted. 
The same message is displayed during calibration, when calibration is not possible or the reading 
is unstable. Clean the measurement cell with replacement of the electrolyte and/or the membrane 
covering the measurement electrodes: If the error indication persists, replace the probe.  
 
Proceed as follows: 
1.  Connect the dissolved Oxygen probe to the instrument. 
2.  Disconnect the temperature only probe, if connected to the instrument, so that the indicated 

temperature is that measured by the dissolved Oxygen probe built-in temperature sensor. 

3.  Turn the instrument on with the ON/OFF key.

 

4.  Wet the sponge contained in the calibrator using 2ml of distilled water. 
5.  Insert the probe into the calibrator. 
6.  Wait at least 15 minutes until the system stabilizes thermally, saturation is reached inside the 

calibrator and the reading is stable. If the membrane and/or electrolyte solution have been 
replaced, wait at least 30 minutes. 

7. Press 

CAL

, then “

F2 = oxy

” (when accessing the calibration, the slope is set automatically to 

1.000 and the previous calibration information is transferred to the “Dissolved Oxygen Probe 
Calibration History” contained in the SICRAM module). The saturation index value, temperature, 
barometric pressure and slope of the probe are displayed from top to bottom. 

 

8.  To proceed with the calibration press 

F3 = CAL

. The calibration value 101.7% and the new 

slope value, corrected according to the new calibration, are displayed. 

9.  If the measurement did not stabilize, press F3 = CAL again to repeat the calibration. 
10. To end the calibration and return to measurement, press 

ESC

The instrument is calibrated and ready for use. 

 

If an error is made during calibration, you can press F1 = RESET to reset the slope value to 1.000. 

Y

ou must repeat the calibration. 

On pressing F3 = CAL during calibration, the instrument checks that the correction to be made 
does not exceed the expected limits for correct functioning. If the calibration is rejected, the display 
shows “

OFS_ERROR”, 

the calibration is ended and the slope value is set to 1.000. 

In case of “

OFS_ERROR

”, replace the electrolyte and the membrane. If the error persists, you can 

try to clean the electrodes using very fine abrasive cloth (better if 

impregnated with ammonia) 

before 

replacing the probe.  

 

Содержание HD 98569

Страница 1: ...ion written in this manual and the instrument that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this manual...

Страница 2: ...2 HD 98569 pH Conductivity Dissolved Oxygen Temperature...

Страница 3: ...t shows a description of the instrument main functions on the display 14 Power supply input 12Vdc for the 5 5mm 2 1mm connector Positive at centre 15 8 pole female MiniDin serial RS232C and USB1 1 2 0...

Страница 4: ...robe can be automatic by recognition of standard solutions 147 S cm 1413 S cm 12880 S cm 111800 S cm or manual with different solutions The dissolved Oxygen probe s quick calibration function guarante...

Страница 5: ...press MEAS EPT DIR The instrument operates in continuous view mode In this mode the displayed measurement is updated every second standard mode EPT MAN The displayed measurement is continuously updat...

Страница 6: ...iption on page 6 6 Dissolved Oxygen It reports the dissolved Oxygen concentration value mg l or saturation index measured by the dissolved Oxygen probe with SICRAM module connected to the central inpu...

Страница 7: ...n description on page 5 It returns the instrument display to the standard measurement mode after pressing the MEM and DISP keys EPT key The key selects the measurement performance mode By pressing the...

Страница 8: ...are present the key allows manual input of the temperature value used for temperature compensation and selection of the unit of measurement C or F See also the description of the EPT key PRINT key It...

Страница 9: ...ge 49 DISP key The feature reports useful information on the instrument and connected probes functioning From top to bottom there are Barometric pressure measured by an internal pressure sensor Temper...

Страница 10: ...leftwards During measurement it decreases the display contrast MENU key Using this button the instrument s menu can be accessed See a detailed description on page 11 Down Arrow In the menu it decreas...

Страница 11: ...irm and return to measurement mode The menu items are listed in this order 1 INFO STATUS HELP 1 1 Instrument Info shows some information on the instrument model types of measurement firmware version s...

Страница 12: ...an be selected only by the administrator see page 17 3 2 4 Clear Log allows to permanently clear all the data stored in the memory using the LOG key Press ENTER to confirm ESC to exit without clearing...

Страница 13: ...for connection using the DeltaLog11 software vers 2 0 and later The instrument waits a connection for 10 minutes then shows an error alternating BT and ERR For the details see the chapter dedicated to...

Страница 14: ...AL blinking message appears the calibration data can still be used The Expired calibration message is indicated in the printout Enter Number of days 0 to disable this feature Note The day is counted a...

Страница 15: ...ONS 6 1 View Oxy History The dissolved Oxygen probe SICRAM module stores the last eight calibrations in the memory This menu item displays each calibration date and slope The Print Oxygen History func...

Страница 16: ...nt can be identified using a code appearing in the printouts and in the stored data This code is managed creation editing erasing using the DeltaLog11 software vers 2 0 and later and this can be perfo...

Страница 17: ...Select the desired user e g User_1 using the arrows and confirm with ENTER 3 Type the old password 00000000 in a new instrument and confirm with ENTER 4 Type the new password not 00000000 and confirm...

Страница 18: ...t partial and total reset Set the user access mode with or without password prompting User Exit Mode Selects the instrument memory management mode Single Memory Mode Set the storage mode when using th...

Страница 19: ...bration performed by the user The instrument stores the last eight pH calibrations performed by the user The last two calibration are stored in the probe s SICRAM memory Upon power on the instrument r...

Страница 20: ...uality of a redox electrode User calibration of the temperature sensor is not required The sensor is calibrated in the factory and the Callendar Van Dusen parameters are recorded in the SICRAM module...

Страница 21: ...1 BUFFER_5 menu items Usually two for the acid one for the neutral and two for the alkaline band will be selected 25 C BUFR_1 NEUTRAL 6 860 7 000 7 648 BUFR_2 ACID 1 679 4 000 4 010 BUFR_3 ALKALINE 9...

Страница 22: ...the pH calibration the previous calibration information are transferred in the pH Electrode Calibration History The current offset and slope values are set to nominal values The offset 0mV the gain s...

Страница 23: ...01 pH 25 C C pH C pH C pH C pH 0 6 98 50 6 83 0 4 01 50 4 06 5 6 95 55 6 83 5 4 00 55 4 07 10 6 92 60 6 84 10 4 00 60 4 09 15 6 90 65 6 85 15 4 00 65 4 10 20 6 88 70 6 85 20 4 00 70 4 13 25 6 86 75 6...

Страница 24: ...olved solids TDS according to the TDS conversion factor which can be modified using the menu the salinity NaCl quantity in the solution expressed in g l By repeatedly pressing F3 cond you can select t...

Страница 25: ...The cell constant K is a piece of information which characterizes the cell It depends on its geometry and is expressed in cm 1 There is no cell capable of measuring the entire conductivity scale accur...

Страница 26: ...ons TDS Coefficient The measurement of total dissolved solids is expressed in g l or mg l Calibration of conductivity The probe calibration can be carried out on one to four points using the standard...

Страница 27: ...ement value with the current calibration 1 326mS in the example The recognized standard calibration solution value Target 1 332 Both values refer to the solution actual temperature that is not compens...

Страница 28: ...f the solution is under 15 C or over 35 C the instrument rejects the calibration and displays NON ADMISSIBLE TEMP Proceed as follows 1 Turn the instrument on with the ON OFF key 2 Set the T temperatur...

Страница 29: ...he next calibration solution is one of the standard solutions automatically detected by the instrument open the menu and re enter the temperature coefficient as it was before the calibration Extract t...

Страница 30: ...e instrument on the other end The SICRAM module contains the probe s information serial number calibration parameters etc The probe together with its cable should be used only with that cable You cann...

Страница 31: ...rement the dissolved Oxygen concentration mg l or the saturation index Immerse the probe in the liquid to a depth of at least 40 mm It is essential that the liquid in contact with the membrane is cont...

Страница 32: ...lution have been replaced wait at least 30 minutes 7 Press CAL then F2 oxy when accessing the calibration the slope is set automatically to 1 000 and the previous calibration information is transferre...

Страница 33: ...embrane Fill the probe s head with the DO 9701 electrolyte solution until the level indicated in the figure filling level Eliminate any possible air bubble in the electrolyte solution Screw the probe...

Страница 34: ...useless consumption of the electrolyte solution Always keep the electrode wet using the special cap provided with the probe and filled with distilled water Polarographic probe and galvanic probe diff...

Страница 35: ...be with SICRAM combined conductivity or dissolved Oxygen probe the reference temperature to compensate all measurements is chosen according to the following principle The temperature only probe has pr...

Страница 36: ...ature variations that are too rapid must be avoided 9 Temperature probes for surface measurements contact probes must be held perpendicular against the surface Apply oil or heat conductive paste betwe...

Страница 37: ...here humidity is high the instrument may be exposed to direct sunlight the instrument may be exposed to a source of high temperature the instrument may be exposed to strong vibrations the instrument m...

Страница 38: ...alkaline second point Generally the pH electrodes have a body in glass handle them with care to avoid breakage In particular the membrane the sensitive element housed in the end part of the electrode...

Страница 39: ...d Check that the standard solutions are in good condition and that the calibration solution temperature is within the range 15 35 C Measurement of the dissolved Oxygen The body of the polarographic pr...

Страница 40: ...MORY Instrument management program error Contact the instrument s supplier and communicate the error message PROBE ERROR A probe with SICRAM module has been inserted when not admissible for that speci...

Страница 41: ...ing and battery voltage falls below the minimum operating level To replace the batteries switch the instrument off and unscrew the two screws on the battery cover counter clockwise After replacing the...

Страница 42: ...llow circuit condensers to discharge completely then reinsert the batteries WARNING ABOUT THE USE OF BATTERIES Batteries should be removed when the instrument is not used for an extended time Flat bat...

Страница 43: ...nd code and CR is the Carriage Return ASCII 0D The XX command characters are exclusively upper case characters Once a correct command is entered the instrument responds with when any wrong combination...

Страница 44: ...ead parameter ID Sample ID 00000001 RK Read cell constant Kcell Chi nominal Kcell 0 700 RL Read parameter Print and Storage Mode Print mem 0 0 only print 1 print and record RM Read parameter Storage M...

Страница 45: ...etting O2 calibration validity number of days 0 999 WP Setting pH Resolution 0 2 decimal positions 1 3 decimal positions WQ Setting salinity for O2 measurement 0 700 0 0 70 0 g l WR Setting conductivi...

Страница 46: ...ble connect the measurement instrument to the first free serial port COM of the PC 3 Turn on the instrument and set the baud rate to 38400 MENU SYSTEM PARAMETERS RS232 Baud_Rate select 38400 using the...

Страница 47: ...take a few minutes 17 The installation procedure is now complete The device will be automatically detected on each new connection In order to check if the entire operation was successful in CONTROL P...

Страница 48: ...plied with its own drivers that should be installed on the PC The HD40 2 printer is fitted with a Bluetooth interface The HD22BT module drivers are contained in the DeltaLog11 CD Rom The instrument pa...

Страница 49: ...y pressing MEM the function keys perform the following MEM F1 STOR It stores the current screen MEM F2 CLR It clears the memory section reserved to the 200 single samples It does not touch the continu...

Страница 50: ...ned in the memory section reserved to single records logged using the MEM key select the screen to be printed using MEM F3 VIEW Print the label using the PRINT key For the details see the previous par...

Страница 51: ...User_2 User_3 or Anonymous SAMPLE ID 00000001 Number of the sample pH EL sernum 01234567 Serial number of the pH electrode pH 7 010 pH measurement pH out of calibration The calibration validity perio...

Страница 52: ...ating security can be ensured only if all the normal safety measures as well as the specifications described in this manual are complied with Its regular functioning and operating security is ensured...

Страница 53: ...the LED located between the and keys stops blinking Now put the liquid container at the center of the base Set the desired rotational speed by pressing the and keys during adjustment the LED located...

Страница 54: ...ABS staff AISI 304 Body ABS base Fe 360 Painting Nickel plated and powder coated Weight 1150g 1900g Bearing surface 100 mm Dimensions Base 215x145x25 mm Staff height 380 mm Height max 450 mm Environm...

Страница 55: ...WD10 12Vdc 1A positive at centre Security of stored data Unlimited Time Date and time Real time Accuracy 1min month max departure Continuous recording LOG key Quantity 9000 samples from the three inpu...

Страница 56: ...ce RS232C USB 8 pole female MiniDIN connector Bluetooth Optional Mains power supply 2 pole connector 5 5mm 2 1mm Positive at centre e g SWD10 Measurement of pH by Instrument Measurement range 9 999 19...

Страница 57: ...nt of total dissolved solids with coefficient TDS 0 5 Resolution Measurement range Kcell 0 01 0 00 1 999mg l 0 005mg l Measurement range Kcell 0 1 0 00 19 99mg l 0 05mg l Measurement range Kcell 1 0 0...

Страница 58: ...e saturation index of dissolved Oxygen Measurement range 0 0 600 0 Resolution 0 1 Instrument accuracy 0 3 1digit in the range 0 0 199 9 1 1digit in the range 200 0 600 0 Setting the salinity Setting D...

Страница 59: ...tion varnishes KP50 0 14pH 0 80 C 3bar Body in glass GEL 1 Teflon ring diaphragm Varnishes cosmetics aqueous emulsions electroplating waters creams deionised water TRIS buffer drinking water stock wat...

Страница 60: ...TRIS buffer waste water viscous samples wine KP70 2 14pH 0 50 C 0 1bar Body in epoxy GEL 1 open hole Dough bread colours varnishes cosmetics creams drinking water aqueous emulsions fruit juices elect...

Страница 61: ...on ring diaphragm Cable L 1m Varnishes cosmetics aqueous emulsions electroplating waters creams deionised water TRIS buffer drinking water fruit juices low ionic content solutions mayonnaise preserved...

Страница 62: ...0 S cm 0 120 C 2 electrode cell AISI 316 Ultra pure water Closed cell measurement Pt100 sensor Cable length 2m SPT01GS K 0 1 0 1 S cm 500 S cm 0 80 C 2 electrode cell in Platinum wire Glass probe mate...

Страница 63: ...ographic probe Silver anode Platinum cathode Galvanic probe Zinc anode Silver cathode Range O2 0 00 60 00mg l 0 00 20 00mg l Operating temperature 0 45 C 0 50 C Accuracy 1 f s 2 f s Membrane Replaceab...

Страница 64: ...Film Air 50 C 250 C 0 3 C TP472I 5 Immersion 50 C 400 C 0 3 C 50 C 300 C 0 6 C 300 C 400 C TP472I 10 Immersion 50 C 400 C 0 3 C 50 C 300 C 0 6 C 300 C 400 C TP49A O Class A Thin Film Immersion 70 C 25...

Страница 65: ...mm paper width four NiMH 1 2V rechargeable batteries SWD10 power supply instruction manual 5 thermal paper rolls HD40 2 The kit includes 24 column portable thermal printer serial and Bluetooth interfa...

Страница 66: ...ode gel filled with screw connector S7 body in glass KP 100 Combined pointed pH electrode flat membrane liquid reference with screw connector S7 body in glass for leather paper ORP ELECTRODES TO BE CO...

Страница 67: ...0 80 C STANDARD CONDUCTIVITY CALIBRATION SOLUTIONS HD8747 Standard calibration solution 0 001 mol l equal to 147 S cm 25 C 200cc HD8714 Standard calibration solution 0 01 mol l equal to 1413 S cm 25 C...

Страница 68: ...0 mm Cable length 2 metres TP474C I Contact probe sensor Pt100 Stem 4 mm length 230 mm contact surface 5mm Cable length 2 metres TP474C 0 Contact probe sensor Pt100 Stem 4 mm length 230 mm contact sur...

Страница 69: ...ard solutions at 147 S cm 1413 S cm 12880 S cm and 111800 S cm 30 Measurement of the dissolved Oxygen 31 How to measure 31 Calibration of the dissolved Oxygen probe 32 Electrolyte solution and or memb...

Страница 70: ...oratory electrode holder with magnetic stirrer 53 HD22 3 Laboratory electrode holder 53 HD 98569 Technical characteristics 55 Technical data in line probes for the HD 98569 instrument 59 2 and 4 elect...

Страница 71: ...tita da una catena di riferibilit che ha origine dalla taratura dei campioni di prima linea dei laboratori accreditati di Delta OHM presso l Istituto Primario Nazionale di Ricerca Metrologica The trac...

Страница 72: ...ive 2002 96 EC the European users of electrical and electronic equipment can return it to the dealer or manufacturer upon purchase of a new one The illegal disposal of electrical and electronic equipm...

Отзывы: