48
T
ECHNICAL DATA OF INSTRUMENT PROBES AND MODULES
Probes for air speed measurement
Hot-wire probe: AP471 S1 - AP471 S2 - AP471 S3 - AP471 S4
AP471 S1 - AP471 S3
AP471 S2
AP471 S4
Type of measurements
Air speed, calculated flow, air temperature
Type of sensor
Speed
NTC Thermistor sensor
Omnidirectional NTC
Thermistor sensor
Temperature
NTC Thermistor sensor
NTC Thermistor sensor
Measuring range
Speed
0.1…40m/s
0.1…5m/s
Temperature
-25…+80°C
-25…+80°C 0…+80°C
Measuring resolution
Speed
0.01m/s (0.1…19.99) - 0.1m/s (20.0 ...)
0.1 km/h
1 ft/min (0…1999) - 10ft/min (2.00·10
3
...)
0.1 mph
0.01 m/s (0.1…5 m/s)
0.1 km/h
1 ft/min
0.1 mph
Temperature
0.1°C (-25…+80°C)
0.1°C (-25…+80°C)
Measuring accuracy
Speed
±0.1 m/s (0.1…0.99 m/s)
±0.05m/s (0.1…0.99 m/s)
±0.3 m/s (1.00…9.99 m/s)
±0.15m/s (1.00…5.00 m/s)
±0.8 m/s (10.00…40.0 m/s)
Temperature
±0.8°C (-10…+80°C)
±0.8°C (-10…+80°C)
Min. speed
0.1 m/s
Air temperature
compensation
0…80°C
Battery life
Approx. 20 hours at 20 m/s using alkaline
batteries
Approx. 30 hours at 5 m/s using
alkaline batteries
Unit of measurement
Speed
m/s – km/h – ft/min – mph
Flow
l/s – m
3
/h – cfm
Duct section for flow
calculation
0.001…1.999 m
2
Cable length
~2m
Содержание DO2003
Страница 2: ...2 HVACR Datalogger DO2003...
Страница 19: ...19 DIMENSIONS 85 87 200 min 880 max 13 8 AP471 S1 110 13 8 AP471 S2 85 87 102 200 min 880 max 13 8 AP471 S3...
Страница 20: ...20 MIN 350 MAX 1020 120 AP471 S4...
Страница 23: ...23 DIMENSIONS AP472 S1 AP472 S2 AP472 S4 Unscrew the handle 3 holding the probe body still in the point 1...
Страница 58: ...58...