Calibrating the voltage input
-
Press the PRG key until the message P9 appears on the display.
-
Press OK; the mV value of the input appears on the display.
-
Simulate a voltage of 0 mV at the input (if the value is between ±25 mV the zero
is calibrated, otherwise the full scale value is calibrated).
-
Using the
▲
and
▼
keys, adjust the voltage value so as to have the correct
voltage value on the display.
-
Press the SET key, the REL symbol lights up on the display indicating that
the instrument is measuring the voltage present at the input using the second
measurement scale.
-
Using the
▲
and
▼
keys, adjust the voltage value so as to have the correct
voltage value on the display.
-
Press the SET key, the REL symbol on the display switches off.
-
Simulate a voltage of 450 mV at the input, corresponding to the full scale value
of the first scale.
-
Using the
▲
and
▼
keys, adjust the voltage value so as to have the correct
voltage value on the display.
-
Simulate a voltage of 1800 mV at the input, corresponding to the full scale
value of the second scale.
-
Using the
▲
and
▼
keys, adjust the voltage value so as to have the correct
voltage value on the display.
-
Press OK to confirm. To quit programming, press PRG repeatedly.
Display
Symbol
description
°C
indicates that the value shown is in °C.
°F
indicates that the value shown is in °F.
pH
indicates that the unit of the value shown is pH.
mV
Indicates that the unit of the value shown is milli Volts.
A
indicates that the relay A is in closed status.
B
indicates that the relay B is in closed status.
REL
- indicates that the value shown corresponds to the closing
thresholds of the contacts of relay A or B;
-
indicates that the offset of the second voltage measurement
scale is being calibrated.
REL flashing
indicates that the value shown corresponds to the opening
thresholds of the contacts of relay A or B.
Δ
- indicates that the instrument is in the parameter setting phase;
-
indicates that the closing and opening thresholds of relays A and
B are being changed;
-
indicates that the manual compensation temperature is being
changed;
-
indicates that the pH electrode is being calibrated.
Error signals
OFL
- Warning which appears during measurement when the value to be
displayed is out of scale.
E1
- Error warning which appears during pH calibration to indicate that the
offset value of the electrode is too high in absolute value.
E2
- Error warning which appears during pH calibration to indicate that the
difference between the mV readings given by the two buffer solutions used
for calibration is too great.
E3
- Error warning which appears during pH calibration to indicate that the mV
readings given by the two buffer solutions used for calibration are too close
(about 50 mV at 25°C).
E4
- Reading error on the EEPROM.
E5
- Error warning indicating that the slope calculation gives a value 20% lower
than the nominal value or gives a negative value.
E6
- Error warning indicating that the slope calculation gives a value 150%
lower than the nominal value.
Order code
DO 9403T-R1:
pH transmitter 4÷20 mA passive or active, power supply 24 Vac,
120x80x56 mm for use in the field.
HD 882 M100/300:
Temperature probe with Pt100 sensor, miniature head, shaft
Ø
6x300 mm.
HD 882 DM100/300:
Temperature probe with Pt100 sensor, DIN B head, shaft
Ø
6x300 mm.
HD 8642:
Buffer solution 4.01 pH.
HD 8672:
Buffer solution 6.86 pH.
HD 8692:
Buffer solution 9.18 pH.
HD 220:
Buffer solution redox 220 mV
HD 468:
Buffer solution redox 468 mV
CP5:
Extension cable. Connector S/wire - TERMINAL BOARD.
CP5/10:
Extension cable L=10m.Connector BNC/S7.
CP5S:
Extension cable L=5m. Connector BNC/S7.
CP5S/10:
Extension cable L=10m. Connector BNC/S7.
KPI 10:
Combined industrial electrode, S7 brass 1“ connector, glass body, Ag/
AgCl sat KCI, Ø 12x120, temperature 0÷130°C, porous Teflon fitting.
KPI 11:
Combined industrial electrode, S7 brass 1“ connector, Rytron body, Ag/
AgCl sat KCI, temperature 0÷100°C, porous Teflon fitting.
KPI 12:
Platinum electrode for Redox measurement,S PG13,5 connector,
pressure 6 bar.
KPI 13:
Platinum electrode for Redox measurement, Rytron body, S PG13,5
connector, Ag/AgCl sat KCl.
KPI GB 210:
Electrode for biotechnology, S7 PG13.5 connector, glass body, gel,
Ø 12x120, temperature 0÷135°C, max 10 bar.
PH-4
PH-5