SM3300 INTERFACES
DELTA ELEKTRONIKA BV
March 2014
SPECIFICATIONS
Page 1 - 3
Isolated Contacts
Specifications
Typical Applications
• Trigger an external safety alarm
• Interact in automated processes
• Switch the output On/Off with a remote 24Vdc signal
• Using a floating signal for triggering the Interlock function
Relay contacts 1... 4
Contact voltage
Contact current
Maximum switching capacity
60 V
2 A
60 W
Floating Interlock
Open circuit voltage
5 V
Floating Enable
Nominal input voltage
Input voltage range
Input impedance
24VDC
15 - 30VDC
12kOhm
Insulation
prog. connectors - internal circuits
prog. connectors - earth
1000 VDC Reinforced isolation acc. EN60950-1 / EN61010-1
max. 60VDC
Safety
EN 60950 / EN 61010
Operating Temperature
−
20 to + 50 °C
Humidity
max. 95% RH, non condensing, up to 40 °C
max. 75% RH, non condensing, up to 50 °C
Storage temperature
−
40 to + 85 °C
Mounting
Pluggable, SM3300 interface slots 1, 2, 3 and 4.
See paragraph 'Hardware Installation' in this manual.
Maximum 4 per unit.
Programming connector
Relay 1 & 2, via a 6 pole push wire or so-called push in connector.
Relay 3 & 4, via a 6 pole push wire / push in connector.
Interlock and Enable via a 3 pole push wire / push in connector.
For all 3 connectors there's a contra header supplied.
Weight
0.14 kg