Delfin ElastiMeter Скачать руководство пользователя страница 6

ElastiMeter User Manual version 1.6 

15.8.2018

 

 

 

O

PERATING INSTRUCTIONS

 

 
 

1.

 

Switch the power on by pressing the yellow operating button. The 
message 

WAIT

 is displayed. 

o

 

The ElastiMeter is ready to measure when the message 

READY

 

is displayed with a short tone. 

 

2.

 

Place the probe perpendicularly, slightly and shortly on the skin, only 
for about 0.5 seconds. 

o

 

The indenter touches the skin first, then the base plate.  

 
 
 
 
 
 
 
 
 
 
 

 

o

 

The indenter imposes a constant deformation when the base 

plate is in full contact with the skin. 

o

 

The force sensors recognize the skin contact and the 

measurement starts automatically. 

o

 

If necessary, support your hand to get a proper contact. 

o

 

If the measurement was correctly performed, the message 

MEASURE sample 2/5

” is displayed. 

 

 

 

 

 

o

 

If the measurement was not correctly done, the message 

USER 

ERROR

 is displayed and the following messages alter.  

 

 

 

 

 

Base plate 

Indenter  

Skin 

Содержание ElastiMeter

Страница 1: ...www delfintech com ElastiMeter User Manual...

Страница 2: ...FACE DESCRIPTION 4 OPERATING INSTRUCTIONS 5 ADJUSTING SETTINGS 7 RECOMMENDATIONS FOR MEASURING WITH THE ELASTIMETER 8 MAINTENANCE CLEANING AND DISINFECTING 9 CALIBRATION 10 TECHNICAL DATA 11 APPENDIX...

Страница 3: ...2...

Страница 4: ...and thus the skin s response under a short term load indicates its instant elastic properties APPLICATIONS Instant skin elasticity measurements give important information on the biophysical propertie...

Страница 5: ...indicator 3 Battery voltage indicator 4 Display 5 Number of the sample 6 Instant skin elasticity ISE value 7 Past values ElastiMeter Accessories available ElastiMeter measuring unit DMC data collectio...

Страница 6: ...5 seconds o The indenter touches the skin first then the base plate o The indenter imposes a constant deformation when the base plate is in full contact with the skin o The force sensors recognize the...

Страница 7: ...ISE reading is the mean of 5 consecutive measurements Readings under the recent value are means of previous measurements If the probe is left on the skin for too long the message OFF SKIN is displayed...

Страница 8: ...stiMeter on by pressing the operating button briefly 2 When the device is on press the button again and hold it until the MENU is displayed Then release 3 The indicator starts to scroll downwards thro...

Страница 9: ...o Avoid skin sites over superficial bones tendons and large veins Environmental conditions The ElastiMeter can be used in normal room temperature Recommended operating conditions are mentioned in the...

Страница 10: ...d with ethanol Set the moistened soft paper on the table and wipe the measurement probe slightly to the paper Use only soft paper tissue with no lint The probe head must be completely dry before use N...

Страница 11: ...Do not use rechargeable batteries 4 Replace and screw the bottom cover in place Do not touch the circuit board with any object CALIBRATION Delfin ElastiMeter is recommended to be calibrated every oth...

Страница 12: ...x 45 x 35 mm o Weight with batteries 153 g o Protection class IP52 o Measurement completion time ISE 0 5 seconds o Measurement repeat cycle ISE 10 measurements per minute typically o Measured paramet...

Страница 13: ...Storage at temperature 5 C to 35 C o Storage at relative humidity 5 to 80 Warranty o Limited two year warranty see Appendix C Bluetooth module certifications o Contains FCC ID QOQBLE113 o Contains IC...

Страница 14: ...s not correct or the probe is left on the skin for too long LO FORCE Applied contact force too low HI FORCE Applied contact force too high OK FORCE Applied contact force correct LO SPEED Skin contact...

Страница 15: ......

Страница 16: ...contact Make sure the measuring surface base plate is fully and evenly in contact with the skin Make sure your hand is supported during the measurement and the base plate touches the skin perpendicula...

Страница 17: ......

Страница 18: ...not apply to any defect failure or damage caused by improper use or inadequate maintenance and care Delfin Technologies shall not be obligated to furnish service under this guarantee 1 to repair dama...

Страница 19: ...Delfin Technologies Ltd info delfintech com www delfintech com...