User related factors 

o

 

Support your hand during the measurement. 

o

 

Make  sure  the  probe  touches  the  skin  perpendicularly  and  the 

measurement site is either flat or convex, not concave.  

 

 

 

 

 
 
 
 
 
 
 

M

AINTENANCE

,

 CLEANING AND DISINFECTING

 

 
The  ElastiMeter  should  be  located  and  stored  in  a  clean,  dry  place  at  room 
temperature,  away  from  extremes  of  draft  and  vibration.  Handle  the  device 
gently. Thumping and shaking may damage the force sensors. 

 

Clean the cover with a mild detergent solution. The measurement probe can be 

carefully  cleaned  with  soft  paper  tissue  moistened  with  ethanol.  Set  the 

moistened soft paper on the table and wipe the measurement probe slightly to 

the  paper.  Use  only  soft  paper  tissue  with  no  lint.  The  probe  head  must  be 

completely dry before use. Never place the device or the probe in contact with 

water or any other liquid. 

 

Batteries 

 

A  set  of  two  1.5  V  AA  alkaline  batteries  lasts  for  several  hundreds  of 

measurements. If the device is not used for several months it is recommended 

to remove the batteries. 

 

The batteries are tested automatically and a “

Low battery

” warning appears on 

the  display  when  the  battery  voltage  is  low.  Change  the  batteries  when  the 

remaining charge in the  battery indicator turns red. If the message “

Replace 

batteries

” appears when the ElastiMeter is turned on, measurements cannot be 

done before the batteries are changed. 

Содержание ElastiMeter

Страница 1: ...www delfintech com ElastiMeter User Manual...

Страница 2: ...FACE DESCRIPTION 4 OPERATING INSTRUCTIONS 5 ADJUSTING SETTINGS 7 RECOMMENDATIONS FOR MEASURING WITH THE ELASTIMETER 8 MAINTENANCE CLEANING AND DISINFECTING 9 CALIBRATION 10 TECHNICAL DATA 11 APPENDIX...

Страница 3: ...2...

Страница 4: ...and thus the skin s response under a short term load indicates its instant elastic properties APPLICATIONS Instant skin elasticity measurements give important information on the biophysical propertie...

Страница 5: ...indicator 3 Battery voltage indicator 4 Display 5 Number of the sample 6 Instant skin elasticity ISE value 7 Past values ElastiMeter Accessories available ElastiMeter measuring unit DMC data collectio...

Страница 6: ...5 seconds o The indenter touches the skin first then the base plate o The indenter imposes a constant deformation when the base plate is in full contact with the skin o The force sensors recognize the...

Страница 7: ...ISE reading is the mean of 5 consecutive measurements Readings under the recent value are means of previous measurements If the probe is left on the skin for too long the message OFF SKIN is displayed...

Страница 8: ...stiMeter on by pressing the operating button briefly 2 When the device is on press the button again and hold it until the MENU is displayed Then release 3 The indicator starts to scroll downwards thro...

Страница 9: ...o Avoid skin sites over superficial bones tendons and large veins Environmental conditions The ElastiMeter can be used in normal room temperature Recommended operating conditions are mentioned in the...

Страница 10: ...d with ethanol Set the moistened soft paper on the table and wipe the measurement probe slightly to the paper Use only soft paper tissue with no lint The probe head must be completely dry before use N...

Страница 11: ...Do not use rechargeable batteries 4 Replace and screw the bottom cover in place Do not touch the circuit board with any object CALIBRATION Delfin ElastiMeter is recommended to be calibrated every oth...

Страница 12: ...x 45 x 35 mm o Weight with batteries 153 g o Protection class IP52 o Measurement completion time ISE 0 5 seconds o Measurement repeat cycle ISE 10 measurements per minute typically o Measured paramet...

Страница 13: ...Storage at temperature 5 C to 35 C o Storage at relative humidity 5 to 80 Warranty o Limited two year warranty see Appendix C Bluetooth module certifications o Contains FCC ID QOQBLE113 o Contains IC...

Страница 14: ...s not correct or the probe is left on the skin for too long LO FORCE Applied contact force too low HI FORCE Applied contact force too high OK FORCE Applied contact force correct LO SPEED Skin contact...

Страница 15: ......

Страница 16: ...contact Make sure the measuring surface base plate is fully and evenly in contact with the skin Make sure your hand is supported during the measurement and the base plate touches the skin perpendicula...

Страница 17: ......

Страница 18: ...not apply to any defect failure or damage caused by improper use or inadequate maintenance and care Delfin Technologies shall not be obligated to furnish service under this guarantee 1 to repair dama...

Страница 19: ...Delfin Technologies Ltd info delfintech com www delfintech com...

Отзывы: