6.3
Basic Setup
The following parameters may be changed from the Basic Setup panel:
•
Field ID: Changing the Field ID is only necessary, if more than one device is mounted within the
Radio Range of the transponders. This requires the field ID to be unique.
•
Reading range control: This will adjust the read range from 10% to 100%
•
Wait for guard timeout (ms): This is the time it takes for the locator to respond when a “Need
Guard” transponder enters the field without its “Guard”.
•
Signal not polled: If this option is activated, the locator signals by flashing yellow when it is not polled.
6.4
Brightness
The brightness of the individual LED colours can be adjusted. LED test can be used to check (first click on
test, then the color).
6.5
Trigger Setup
•
Use input 1 of locator as trigger: The locator field can be triggered by an external signal on input
1.
•
Trigger time after edge (ms): For edge controlled triggering, the time has to be specified.