Controlling the PDG-2500 from a computer
1.
Disable the PDG-2500
Send the command “OUTPUT:DISABLE”.
2.
Set the pulse trigger frequency
Send the command “FREQ n”, where n = [5 to 1000000]. Units are hertz.
3.
Set the pulse duty cycle
Send the command “DUTYCYCLE n”, where n = [0.01 to 100.00].
Units are percent.
4.
Set the analog voltage
Send the command “ANALOG:SETPOINT n”, where n = [0 to 65536].
Units are DAC counts of 5 V/65535 = 76.293
µ
V each.
5.
Enable the PDG-2500
Send the command “OUTPUT:ENABLE”.