Set the Analog Output
ANALOG:SETPOINT?
ANALOG:SETPOINT n
1. On the Home screen, press the
Analog Output
icon.
2. The
Set Analog Output
screen appears.
3. Press the encoder knob to select a resolution: x1,
x10, x100, x1,000, or x10,000.
4. Turn the encoder knob to change the Analog Set
point: 0 to 65,535. The larger the DAC count the
higher the V
out
set point. Both the DAC count and the
voltage are displayed.
5. On the screen, press the
Done
icon.
Set the Single Shot
SINGLE:SHOT?
SINGLE:SHOT n
1. On the Home screen, press the
Frequency
icon.
2. The
Set Frequency
screen appears. Press the
Single Shot
check box.
3. The home screen appears in single shot control.
4. To change the single shot pulse width, press the
Single Shot
icon.
5. Press the encoder knob to select a resolution: x80,
x160, x1,000, x10,000, x100,000, x1,000,000,
x10,000,000, or x100,000,000 ns.
6. Turn the encoder knob to change the single shot: 80
ns to 1.0000000 second.
7. On the screen, press the
Done
icon.
8. Press
Enable
icon to turn on the single shot pulse for
the set pulse width.
9. To change back to duty cycle control, press the
Duty
Cycle Control
check box.