Define Instruments TM-USB Скачать руководство пользователя страница 2

2

C

Click 

Browse...

 to change the installation location (if 

required), and then click 

Next

.

D

Click 

Next

 to begin installing. This may take a few 

minutes.

E

Click 

Close

 to exit the 

TM Setup

 wizard.

Содержание TM-USB

Страница 1: ...nstall file from our website www defineinstruments com tmsetup Windows 2000 XP Double click TMInstall msi Windows Vista 7 Right click setup exe and select Run as administrator B The following dialogue will appear Click Next to begin setup ThankyouforpurchasingaTM USBreusablecommunications kit This booklet contains 1 TM Setup installation instructions p1 2 TM USB connection instructions p3 3 TM Set...

Страница 2: ...2 C Click Browse to change the installation location if required and then click Next D Click Next to begin installing This may take a few minutes E Click Close to exit the TM Setup wizard ...

Страница 3: ...r as shown below 2 Always use the windows Safely remove hardware function before unplugging from your PC 3 Please install TM Setup p1 3 before you connect your USB programmer to your PC After you have installed the TM Setup software the TM USB programmer can be connected between the transmitter and your PC USB port TM USB Programmer Transmitter Interface cable USB Extension Cable TM USB kit contai...

Страница 4: ... Setup installer will automatically place this icon on your desktop Double click to get started B Connect your TM transmitter to your PC via the USB programmer and interface cable and then click Connect C The following dialogue will appear while the software establishes a connection with the transmitter The transmitter s settings will then be displayed ...

Страница 5: ... transmitter from the TM USB the connection will time out and close automatically If you connect another transmitter before the time out period expires the settings will be read out of the new transmitter and communications will start again F MODEL DETAILS Panel shows transmitter name software version serial no and status Status Values OK Input and Output OK NO LOOP No 4 20mA output connected SENS...

Страница 6: ...n The following dialogue will be displayed during programming When programming is complete the Test and Wiring panels will be updated to reflect the new changes Neverdisconnectthetransmitterduringprogramming If you do so the transmitter will be left in an unknown state and may not operate See Appendix A for information on errors other dialogues that may appear during programming H TEST This panel ...

Страница 7: ...nufacturer This will upset factory output calibration and will not influence the test panel I COMMS This panel shows TX and RX message counts indi cating good communica tions with the unit J INPUT ZERO ADJUST This panel allows you to apply an offset calibration value to the input Click Calibrate to begin calibration The following dialogue will appear Use a positive value to increase the input read...

Страница 8: ... Documents TM Setup cal serialnumber txt If you change the input type i e from RTD to T C the calibration value will be cleared Changing the range scaling does not clear the calibration See Appendix B for information on errors other dialogues that may appear during zero adjustment K HELP INFORMATION Transmitter Installation Guide PDF manual for the attached transmitter or a list of manuals if no t...

Страница 9: ...type a custom file name if desired M AUTO REPEAT PROGRAMMING Once a txcal file has been saved see step L above it can be used to auto program transmitters to the same settings as follows A Close TM Setup program B Attach new TM transmitter to the computer C Double click on the txcal file TM Setup will auto start connect program the attached transmitter and then close again Multiple transmitters ca...

Страница 10: ... connections required for the selected input The wiring panel will be updated toreflectchanges when you click Program see step G Please also refer to the transmitter manuals for wiring information See www defineinstruments com manuals ...

Страница 11: ...appear and the software will terminate connection to the transmitter Connect the transmitter to the computer again ensuring that all cables and adaptors are fully pushed in Then click Connect and repeat the programming procedure B INPUT ZERO ADJUST VERSION COMPATIBILITY This feature requires transmitter firmware V 1 20 or later as shown in the Model Details panel If the transmitter does not have t...

Страница 12: ...12 12 TM USB MAN 11V06 1118 PH 64 9 835 1550 FAX 64 9 835 1250 AUS 1800 810 820 defineinstruments com ...