Calibration Memory

When measuring a particular part with a specific powder, it is often
convenient to be able to store any unique calibration adjustments made
for that application. Then, if you return to that part, the corresponding
Cal setting can be conveniently and quickly restored.

A Cal setting is any of the three calibration adjustments (pg. 8). The
PosiTector PC always displays the current calibration setting (ex. Cal
3) in the upper right corner of the display. 

The gage reverts back to its factory default Cal 1 setting after a Reset
(pg. 13) or if the user deletes all calibration settings (below).

-Creates a new calibration setting using the next available number
(Maximum of 9). By default, these new Cal settings are initially created
with the gage's factory settings as indicated with the       symbol that
appears at the bottom of the display.

-Loads an existing calibration setting. Use the (-) or (+) buttons to scroll
the List box until the desired setting appears, then press 

(.:::.

-Removes a calibration setting completely from the list. That Cal
number can be reused later with the "New" command.

-Sends a list of all Calibration Settings to the IR printer using the gage's
built-in IR port.

Cal Memory

New

Open

Delete

Print

Содержание PosiTector PC

Страница 1: ......

Страница 2: ...hout this manual the symbol indicates more information about the particular topic or feature is available on our website Go to www defelsko com manuals Power up Power down Step 1 Power up the probe by...

Страница 3: ...m height 4 Release the button when you hear a double BEEP or see a thickness measurement on either display Both units will power down automatically after approximately 5 minutes of inactivity Measurem...

Страница 4: ...ration Setting Unit of Measurement Maximum Amplitude Dot Probe Display button not pressed Amplitude Indicator Results Indicator Distance Target Bar Probe Batteries Status pg 15 Bluetooth Function Indi...

Страница 5: ...ltrasonic echo A high position indicates good probe orientation relative to perpendicular The bar will generally reach higher positions when measuring thin powder and when the probe is at a 90 degree...

Страница 6: ...Statistics Zero Cal Settings Setup Exit Appears when probe is ON Current selection is displayed with darkened background To navigate press to scroll DOWN to scroll UP and to SELECT To exit any menu pr...

Страница 7: ...as the PosiTector 6000 Compare results A successful verification requires that the average of a series of measurements by both instruments be within their combined tolerances If not adjust the PosiTec...

Страница 8: ...Adjustment is the act of aligning the gage s predictions to match those of a known cured sample Begin by performing a Verification of Accuracy pg 6 to determine if there is a need for an adjustment If...

Страница 9: ...tolerances the PosiTector PC can be adjusted using one of three methods 1 1 point Adjustment The first and most common choice 2 2 point Adjustment If a 1 Pt Adjust does not improve accuracy 3 Powder...

Страница 10: ...ess with a magnetic or eddy current coating thickness gage Record the results 10 Select 2 Pt Adjust then select the appropriate batch names when asked 11 Adjust the displayed batch averages lower or h...

Страница 11: ...y default Cal 1 setting after a Reset pg 13 or if the user deletes all calibration settings below Creates a new calibration setting using the next available number Maximum of 9 By default these new Ca...

Страница 12: ...d each measurement will now be simultaneously shown on the display and stored into this new batch On screen statistics are immediately updated with each measurement New batch names are date stamped at...

Страница 13: ...ss both and buttons simultaneously to cancel printing Downloading Measurements A computer can view and download measurements stored in memory using the included USB cable Measurements are not erased f...

Страница 14: ...t symbol appears It performs the same function as a menu Reset with the addition of Units microns Language English and Flip Display Normal A probe Zero pg 6 must be performed after any Reset This opti...

Страница 15: ...e USB Drive A formatted HTML report is viewed by selecting the index html file found in the root directory Optionally readings0 txt files located in each batch folder i e B1 provide access to measurem...

Страница 16: ...ed from the screen by no more than 0 2 MPa 20 30 psi of compressed air Do not direct air directly into the probe Replacement screens are available Changing the Batteries The battery symbol displays a...

Страница 17: ...o com support Technical Data Conforms to ASTM D7378 Base unit dimensions 146 x 64 x 31 mm 5 75 x 2 5 x 1 2 Probe dimensions 222 x 50mm diameter 8 75 x 2 diameter Alkaline Batteries Base unit 3 AAA bat...

Страница 18: ...Warning The PosiTector PC Powder Checker contains a very powerful magnetic mount for attaching the protective rubber holster to nearby steel structures As with all magnets avoid contact with electroni...

Страница 19: ......

Страница 20: ......

Отзывы: