Coating

Substrate

Probe

Couplant

Measurement of dense (hard) coatings.

A significant echo occurs at the probe/coating
interface. The relative strength of this echo
compared to the coating/substrate echo requires that
Gate 

A

be increased.

Measurement of total thickness in multi-layer
coating applications.

Multiple coating/coating/substrate interfaces may
generate several echoes. The user may need to
adjust Gates to ignore echoes from coating/coating
interfaces.  

limits of the gage and assumes the largest “echo” is
the coating/substrate echo (single coating
applications only). Adjustable measurement Gates
have been provided for the user to force the
instrument to ignore echoes from unwanted
boundaries within the sample. Several examples
below help to illustrate the use of Gates for specific
applications.

Measurement of coatings with rough surfaces.

When rough coatings are measured, the gage
typically identifies the thickness from the top of the
coating peaks down to the substrate (#1).  Couplant
fills the voids between the probe and the coating (#2)
creating an additional interface.  If echoes from the
couplant/coating interface (#2) are stronger than the
coating/substrate interface (#1), an adjustment
(increase) of Gate 

A

may be required for the gage to

display the weaker (#1) echo.

Содержание PosiTector 200

Страница 1: ...PosiTector 200 INSTRUCTION MANUAL v 2 0 Simple Durable Accurate Ultrasonic Coating Thickness Gage Standard and Advanced...

Страница 2: ...ve ultrasonic principle to measure coating thickness on a wide variety of substrates The PosiTector 200 probe emits a high frequency sound pulse that travels into the coating via a coupling gel and re...

Страница 3: ...to a known thickness if necessary see pg 8 4 MEASURE the part see pg 5 Couplant Couplant Couplant is required to propagate ultrasound into the coating Water is a good couplant for smooth coatings Use...

Страница 4: ...enu or select Exit from the Menu List boxes have a down arrow on the right hand side Use the and buttons until your desired choice appears then press to select this choice and move focus onto the next...

Страница 5: ...to the protective case There is no need to disconnect the probe from the Gage during storage How to Measure How to Measure Zero Menu Option Zero Menu Option Zero The probe must be periodically zeroed...

Страница 6: ...alues do not have to be adjusted But some conditions may exist in the user s application that cause the Gage to display very low or non repeatable readings These conditions include rough or textured c...

Страница 7: ...o accept Gate values and exit Use the Cursor to measure multi layer coating systems with more than 3 distinct layers NOTE Here are some typical Gate settings Expected paint thickness Gate A Gate B 500...

Страница 8: ...of couplant onto the reference stan dard 2 Measure the reference standard 3 Lift the probe Select the Adj Thickness menu option 4 Adjust the display down or up to the reference standard thickness 5 Pr...

Страница 9: ...nd made the current batch Statistics are displayed and each measurement will now be simultaneously shown on the display and stored into this new batch On screen statistics are immediately updated with...

Страница 10: ...a and readings can be exported to a document or spreadsheet Remove the last reading from the current open batch by pressing NOTE View Delete removes a batch completely from memory The name is deleted...

Страница 11: ...anced models only A more thorough Reset can be performed by holding the button when the Gage is powered down until the Reset symbol appears It performs the same function as a menu Reset with addition...

Страница 12: ...ating layer thickness For example if the number of layers is set to 3 the 3 highest peaks between the A B Gates are selected as the interfaces between these layers The peaks that the Gage selected are...

Страница 13: ...ayers Measuring Multiple Layers Advanced models only Advanced models of the PosiTector 200 are capable of numerically displaying up to 3 individual layer thicknesses in a multi layer system To setup t...

Страница 14: ...within the combined accuracy of the Gage and the reference standards Adjustment Adjustment or Calibration Adjustment is the act of aligning the Gage s thickness readings to match that of a known refe...

Страница 15: ...LL reflections within the measurement Application Notes Application Notes The PosiTector 200 uses an ultrasonic principle to measure coating thickness of most coatings on most substrates An ultrasonic...

Страница 16: ...te echo single coating applications only Adjustable measurement Gates have been provided for the user to force the instrument to ignore echoes from unwanted boundaries within the sample Several exampl...

Страница 17: ...Gate A See pg 6 Gage readings are much higher than expected Gage may be measuring both the coating and substrate Lower the value of Gate B See pg 6 Gage does not yield accurate or consistent results S...

Страница 18: ...ress Website www defelsko com support Changing The Batteries Changing The Batteries The battery icon displays a full bar with fresh alkaline batteries installed As the batteries weaken the bar will be...

Страница 19: ...rmined in our sole discretion that the returned instrument is defective as to workmanship or material the instrument will be repaired or replaced at DeFelsko s sole option In no event shall DeFelsko b...

Страница 20: ...ten permission from DeFelsko Corporation DeFelsko PosiTector and PosiSoft are trademarks of DeFelsko Corporation regis tered in the U S and in other countries Other brand or product names are trade ma...

Отзывы: