Technical Specifications and Operating Procedure. PILATUS3 S 6M-NSRRC Detector System. Version: V1.0 

20/33 

5.

 

Installing the Detector System 

5.1.  Carrying 

The  detector  head  has  four  transport  hooks,  which  have  to  be  used  for  lifting  the  detector  head.  For 
tilting/rotating the detector head, two hooks can be used: Lift the detector head up so it hangs diagonally, 
then set it down towards the other side. 

Only use an appropriate lifting device with a carrying capacity of at least 1000 kg. 

Make sure 2 persons always move the detector head. 

When using the transport hooks, make sure they are properly tightened. 

When moving the detector head, make sure all cables and hoses are disconnected. 

 

The detector electronics has a transport hook, which has to be used for lifting the detector electronics. 

Only use an appropriate lifting device with a carrying capacity of at least 100 kg. 

Make sure 2 persons always move the detector electronics. 

When using the transport hook, make sure it is properly tightened. 

When moving the detector electronics, make sure that all cables are disconnected. 

 

5.2.  Mounting of the Detector Head 

Install the Mylar

®

 window inside the beamline sample chamber before mounting the detector head. 

Do not touch the Mylar

®

 foil. Contact DECTRIS support in case of contamination of the window. 

 

With  the  blank  flange  still  attached  rotate  the  detector  head  so  the  x-ray  sensor  points  downward  (see 
above  how  to  rotate).  Remove  the  blank  flange  and  mount  the  detector  head  on  top  of  the  sample 
chamber. 

Remove the blank flange in a clean, dust-free and humidity-controlled environment. Minimize the 

time the sensors are exposed without protection. 

BE VERY CAREFUL when installing the detector head or the blank flange and avoid touching the 

sensor under all circumstances, as this will most likely result in a damaged sensor. Make sure the blank 
flange does not tilt inward through the opening at any time. Stay clear of the detector head volume with 
any kind of object (screw-

driver, finger, …). This operation is restricted to trained personnel. Damage 

caused by improper handling is not covered in the warranty. 

After installation immediately flush the chamber with dry air or pump down the system to keep 

humidity as low as possible. 

Содержание PILATUS3 S 6M-NSRRC

Страница 1: ...DECTRIS 5400 Baden Switzerland www dectris com Technical Specifications and Operating Procedure PILATUS3 S 6M NSRRC Detector System Version V1 0...

Страница 2: ......

Страница 3: ...etector Dimensions and Connectors 13 4 1 The PILATUS3 S 6M NSRRC Detector 14 4 1 1 Front Side of the Detector Head 14 4 1 2 Back Side of the Detector Head 15 4 1 3 Front Side of the Detector Electroni...

Страница 4: ...g Procedure PILATUS3 S 6M NSRRC Detector System Version V1 0 4 33 7 1 Startup Procedure 27 7 2 First Commands 27 8 Turning off the Detector 28 9 Storing the Detector 29 10 Cleaning and Maintenance 30...

Страница 5: ...6 PILATUS3 S 6M NSRRC detector electronics viewed from the front left and its connection layout right 16 Figure 7 The PILATUS3 S 6M NSRRC detector electronics viewed from the back 16 Figure 8 Configur...

Страница 6: ...sion V1 0 6 33 1 Document History 1 1 Current document Version Date status prepared checked released 1 0 02 11 2015 draft BL Table 1 Document history 1 2 Changes Version Date Changes 1 0 02 11 2015 Fi...

Страница 7: ...receive the necessary transport and shipping information Make sure that the original packaging is used when returning the system When returning the detector system for repair be sure to fill out and i...

Страница 8: ...ounded Make sure that the PC has adequate ventilation Improper use of the DECTRIS Ltd detector system can compromise safety and its functionality is also no longer guaranteed Do not modify the detecto...

Страница 9: ...Format 2463 x 2527 6 224 001 pixels Area 423 6 x 434 6 mm 2 Intermodule gap x 7 pixels y 17 pixels 8 5 of total area Dynamic range 20 Bits 0 1 048 573 Counter overflow state 1 048 573 Counting rate p...

Страница 10: ...llowable coolant pressure in the cooling circuit is 2 bar Coolant Use mixture of 2 3 distilled water and 1 3 Ethylenglycol Power consumption 580 W Dimensions W x H x D Detector head 836 x 880 x 654 mm...

Страница 11: ...l common supply voltages Detector external trigger input 2 1 V 5 0 V high level 0 0 V 0 8 V low level 50 impedance 5 0 V absolute maximum Applying a higher voltage will damage the input circuit Detect...

Страница 12: ...condensing Table 6 Detector operating conditions Note that the interior humidity under operating conditions must be 25 When storing the detector make sure the temperature and humidity inside the tran...

Страница 13: ...Dimensions and Connectors Figure 1 Rendering of the PILATUS3 S 6M NSRRC detector consisting of detector head left with the front side facing down and the back side facing up and detector electronics r...

Страница 14: ...tion the detector head is flanged directly onto the vacuum system The detector head comes with a blank flange as a protective cover during e g storage transport or vacuum tests bake out Once the blank...

Страница 15: ...TA 12 DATA 11 POWER 4 DATA 20 DATA 19 DATA 18 DATA 17 DATA 16 POWER 5 DATA 25 DATA 24 DATA 23 DATA 22 DATA 21 POWER 6 DATA 30 DATA 29 DATA 28 DATA 27 DATA 26 POWER 7 DATA 35 DATA 34 DATA 33 DATA 32 DA...

Страница 16: ...5 DATA 26 DATA 27 DATA 28 DATA 29 DATA 30 POWER 6 DATA 31 DATA 32 DATA 33 DATA 34 DATA 35 POWER 7 DATA 36 DATA 37 DATA 38 DATA 39 DATA 40 POWER 8 DATA 41 DATA 42 DATA 43 DATA 44 DATA 45 POWER 9 DATA...

Страница 17: ...pproved to EN 60950 1 Cable Power cable with an IEC C13 connector EXT IN External trigger input TTL Use a Lemo Type 00 NIM CAMAC cable EN OUT TTL output signal high when counting is enabled Use a Lemo...

Страница 18: ...TUS Processing Unit PPU 2 x 10 Gbit Ethernet connection High bandwidth connection TCP IP Client local or remote TCP IP Client local or remote Figure 8 Configuration of the PILATUS3 S 6M NSRRC detector...

Страница 19: ...ng type In air or vented vacuum In vacuum Operating temperature The cooling unit has always to be set to a temperature of 23 C Prior to and during pumping down and venting the cooling unit has to be s...

Страница 20: ...tector electronics make sure that all cables are disconnected 5 2 Mounting of the Detector Head Install the Mylar window inside the beamline sample chamber before mounting the detector head Do not tou...

Страница 21: ...d in Figure 9 to bolt the detector down Make sure the mounting brackets are mounted and properly tightened using the four optional M4 screws Do not place the detector near heat sources or in a place s...

Страница 22: ...The detector should be mounted using the four internal threads M6 x 1 The four M6 screws must not intrude into the detector more than 16 mm Make sure the detector is properly mounted It is strictly fo...

Страница 23: ...idity see section 6 for how to measure humidity As an alternative to nitrogen oil free dry air of 20 relative humidity can be used For reliable operation we recommend a dry air of 5 relative humidity...

Страница 24: ...to section 7 and execute the correct startup procedure In vacuum operation The temperature and humidity control cannot prevent condensation issues and resulting damage to the sensor due to improper u...

Страница 25: ...nical Specifications and Operating Procedure PILATUS3 S 6M NSRRC Detector System Version V1 0 25 33 Figure 11 Error message when temperature or humidity is out of range Figure 12 Correct start up mess...

Страница 26: ...unit Connect the PC and the detector data cables Attach a monitor keyboard and mouse to the PC In vacuum operation Mount the detector head properly to the vacuum system Mount the detector electronics...

Страница 27: ...ch initializes the detector system and opens the Camserver and TVX windows If you want to control the detector with a TCP IP client type camonly in the directory p2_det camonly starts a script which i...

Страница 28: ...the back of the detector electronics Keep the detector under vacuum In vacuum operation prior to venting and pumping down Always make sure the detector is SWITCHED OFF and WARMED UP to room temperatu...

Страница 29: ...pump down the detector head vacuum chamber and vent it with nitrogen or dry air Quickly seal the vacuum chamber avoiding as much as possible an exchange between the dry air inside and the humid ambie...

Страница 30: ...enance free Maintenance instructions for the cooling unit are provided in the cooling unit s manual The following procedures have to be done periodically What When Who Check and if necessary clean the...

Страница 31: ...between the PC and the detector Avoid tangling or strong bending of the Ethernet data cables Check configuration of the Ethernet adapters Check the status of the LINK LEDs If PC and detector are powe...

Страница 32: ...itialization Detector is not properly initialized Run the following commands in TVX setdac calibdet expose 1 Check the status of the POWER LED at the back of the detector If it is red and the TEMP LED...

Страница 33: ...Name and phone No Company List of all control settings Describe the problem and check boxes below that apply to the problem Checked all cables Problem on power up Detector system is unstable What pow...

Отзывы: