Technical Specifications and Operating Procedure. PILATUS3 S 6M-NSRRC Detector System. Version: V1.0
17/33
4.1.5. The Status LEDs
LED
Description
EN
Yellow, if the detector is in counting mode.
TEMP
Normally green. Turns red if the detector temperature or
humidity is out of the limits.
POWER
Normally green. Turns red if there is a power failure or if
the detector temperature or humidity is out of the limits.
Table 8: The meaning of the status LEDs on the detector back plane.
4.1.6. Connectors and Connecting Cables/Pipes
Connector
Description
DATA 1 (left, Figure 7)
RJ45 Cat 6 S/FTP cable for data transfer (DATA 2, 3, 4
unused).
POWER
Main voltage 100
– 240 VAC; 50/60 Hz, Fuse 6.3A
UL recognized to UL 60950/CSA 22.2 No. 60950-00, and
TUV approved to EN 60950-1
Cable:
Power cable with an IEC/C13 connector.
EXT IN
External trigger input TTL.
Use a Lemo
®
Type 00 (NIM/CAMAC) cable.
EN OUT
TTL output signal; high when counting is enabled.
Use a Lemo
®
Type 00 (NIM/CAMAC) cable.
Functional ground of the detector system.
Although the detector might be grounded via the
mounting bolts, the detector can be grounded additionally
via the functional ground connector at the back (M4 screw-
in tap hole) to establish a defined grounding.
INTERLOCK
Possibility to prevent high voltage discharges in case of
vacuum leaks
in-air operation: Interlock not necessary, use the shorting-
plug provided
in-vacuum operation: only short-circuit the two contacts,
when the vacuum is in a safe regime (either 1e+3 mbar or
<1e-2 mbar) to protect the detector
Use an interlock circuit, which is electrically
isolated from the detector and simply connects (high
voltage on) or disconnects (high voltage off) the two