background image

 

 Mounting set  

INFM7X02 

__________________________________________________________________________________ 

__________________________________________________________________________________

Sida 6/8 

 

 

Now use the leveling tool that is included in the delivery of Infinity. Since the adapter bar 

have splines inside it will not be possible to get the bubble of the spirit level exactly right. 

Just choose the most correct position. 

 

The leveling tool has two marks “H” and “L”. When the bubble of the spirit level is towards 

the “H” the castors of the wheelchair will lift higher from the ground compared to if the 

bubble of the spirit level is towards the “L”. 

 

Adjustment of the angle can be done in two different manners. Both is descriped bellow.  

 
 

 

Alt1. 

Using the adjustment tool that 

is included in the delivery of 

Infinity (not this mounting set), 

set the angle of the adaptorbar 

so that the spirit level is in level. 

Inside the pipes there are 

splines so when rotating you 

will feel and hear a clicking. 

When desired angle have been 

accomplished, check that the 

center pipe is centered in the 

wheelchair. Now tighten the 

screws in the clamps (C).  

 

 

WARNING!

 

The screws holding the clamps 

(C) shall be tightened to 7-9Nm 

of torque.  

 

 

 

 

Содержание INFM7015

Страница 1: ...Assembly Instructions INFM7015 INFM7115 INFM7215 Infinity Mounting set Invacare Action 5 Rigid Swing Away 2021 10 Revision 1 ...

Страница 2: ... operation that if not correctly performed may lead to material damage Particularly important or explanatory information General instructions These instructions shall always be followed when placing the mounting set INFM7015 INFM7115 INFM7215 To be able to connect Infinity to the wheelchair Infinity also needs to be adjusted according to the assembly instructions that is included in the delivery o...

Страница 3: ..._______________________________ __________________________________________________________________________________ Sida 3 8 2 Overview A Center pipe adapter bar B Outer pipe adapter bar C Clamp adapter bar D Adapter plate E QR Pin F Universal clamp G 19mm clamp A B C F D E G ...

Страница 4: ...adjustment of adaptor First mount the adapter plate F universal clamp F wedge I sleeve J screw K and washer L onto the wheelchair Tighten loosely so that the parts can move along the frame Mount the 19mm clamp G around the frame using screw K washer L and locking nut M and tighten loosely Assemble both sides WARNING Loctite 243 or eq shall be applied to the screws K for the universal clamp F M L K...

Страница 5: ...er pipe B are parallell otherwise it can be hard to assemble it into the holders D It shall be possible to move the center pipe A sideways and also rotate it if correctly assembled First tighten the screws in the 19mm clamp G it shall be possible to move the center pipe sideways and also rotate it even after tightening of the screws Now tighten the screws in the universal clamp F Check that the ar...

Страница 6: ...l is towards the H the castors of the wheelchair will lift higher from the ground compared to if the bubble of the spirit level is towards the L Adjustment of the angle can be done in two different manners Both is descriped bellow Alt1 Using the adjustment tool that is included in the delivery of Infinity not this mounting set set the angle of the adaptorbar so that the spirit level is in level In...

Страница 7: ...e of the adaptorbar Only mount the inner pipe on one side and then rotate the center pipe to the correct angle Now mount the other inne pipe in the same position angle as the first one Check that the center pipe is centered in the wheelchair Now tighten the screws in the clamps C WARNING The screws holding the clamps C shall be tightened to 7 9Nm of torque Now read the assembly instructions that w...

Страница 8: ...________________________________________________ __________________________________________________________________________________ Sida 8 8 Decon Wheel AB Södra Ekeryd 119 SE 314 93 Hyltebruk Sweden Tel 46 0 345 40880 Fax 46 0 345 40895 E mail info decon se ...

Отзывы: