Adjustments
17
2. Connect the ac line adapter to the radiance meter and plug it in to a
wall outlet.
NOTE
Make sure that your radiance meter is in calibration.
3. Turn on the power switch (down position).
NOTE
Do not use the POWER ON W/BACKLIGHT position when you
are using the radiance meter with batteries.
4. Set the range switch to the second position from the top (1.999).
5. Place the occluder firmly against the center of the screen. Take a red
china pencil from the color alignment kit and draw an arc or circle
around the occluder.
NOTE
You must take all meter readings with the meter centered in this
marked arc or circle .
M A - X 0 5 7 3 - 8 8