- 9 -
automatically adjusted to the higher value by the system software if selected Fluence is
too high and can not be achieved with currently selected
Pulse
Duration.
Note: Pulse
Duration
settings range is always limited by selected Fluence, i.e. some
pulse widths may not be available at certain combinations of the Fluence and Beam Size
settings. This is normal.
When
Save/Ready
button is pressed after parameters selection and editing is finished,
new set of operating parameters is saved for the current treatment session and the system
switches to
Ready
mode.
LONG PULSE MODE
All pulses with Durations above 10 ms (Nd:YAG) and 7 ms (Alexandrite) are
“stretched”, i.e. consist of a train of shorter pulses spaced in time. The sum of the pulse
widths in each train is equal to 10 or 7 ms respectively. This mode of operation is called
Long Pulse Mode and is designed to provide a certain degree of pain relief for patients
with too sensitive skin. It is not recommended to use this mode for treatment of areas
with very fine (thin) hair.
Example:
To select and then adjust the
Fluence
setting for the patient with Skin Type 3
and coarse dark hair for underarm hair removal, do the following:
1.
Press SKIN TYPE button and then select the Skin Type III option from the
horizontal row (see Fig. 5-4)
2.
Press HAIR TYPE button and select Coarse from two options that will appear on
the screen.
3.
Press HAIR COLOR button and select III (the dark one) from three options in the
horizontal row.
4.
Press BODY AREA button and select Underarms from the options in the
horizontal row or use the body figure on the left and pick the underarm area using
the touch screen.
5.
If there is a desire to adjust the recommended settings recalled from the system
memory, use
>
button if you want to increase the Fluence and
<
button if the
Fluence must be decreased or use a slide bar to quickly change its value.
6.
When the desired Fluence value is displayed, adjust the other parameters (if
necessary) and then press
Save/Ready
button to switch the laser to
Ready
mode.
Содержание POLYLASE MX
Страница 6: ...2 Figure 1 1 POLYLASE MX Dual Output Laser...
Страница 27: ...9 Figure 3 7 Front and side of the POLYLASE MX Laser...
Страница 30: ...12 Main Control Panel Figure 3 9 Control Panel Parameter Setting Edit Screen...
Страница 33: ...15 Figure 3 10 Labels front of the POLYLASE MX Laser Figure 3 11 Labels rear of the POLYLASE MX Laser...
Страница 34: ...16 Figure 3 12 Handpiece Label Laser Aperture Label 4000 001A...