3.a.5.c S
ETTING THE
M
ODEL
3540’
S
F
REQUENCY
I
NPUT
R
ANGE AND
S
ENSITIVITY
The
Model 3540 Frequency Conditioner
must be set to match the full-scale
input range of the frequency source.
1.
If, after pressing the
RANGE
button, you want to display the current input-
range setting, press
ENTER
. Otherwise, press
RANGE
again and proceed to
Step 5. If you press
ENTER
, the unit will display a number representing the
current input-range setting (in Hz): "250," "500," "1000," "2000," "4000," "8000,"
"16000," or "32000."
2.
To change the input range, you can cycle forwards or backwards through the
allowed numbers by pressing the top or bottom segment, respectively, of any
digit display (even if it is blank).
3.
When the desired frequency range appears, press
ENTER
. The unit will dis-
play "SENST" (for "SENSITIVITY"):
The "input sensitivity" is the minimum amplitude of AC input that will trigger an
output—that is, the lowest-valued input signal for which a valid reading is
required.
4.
If you want to display the current sensitivity setting, press
ENTER
. Otherwise,
press
RANGE
again and proceed to Step 6. If, for example, the conditioner is
set for a sensitivity of 1 V to 20 V, when you press
ENTER
it will display "1."
5.
To change the input sensitivity, you can cycle forwards or backwards through
the allowed settings (".1" for .1 V to 5 V; "1" for 1 V to 20 V; and "10" for 10 V to
100 V) by pressing the top or bottom segment, respectively, of any digit dis-
play (even if it is blank).
6.
When the desired sensitivity appears, press
ENTER
to return to the "RANGE"
display. Then press
SETUP
to return to the "Setup" display, and proceed to
Section 3.a.6.
3.a.5.d S
ETTING THE
M
ODEL
3560’
S
V
OLTAGE
I
NPUT
R
ANGE
The
Model 3560 Voltage Conditioner
must be set to match the full-scale input
range of the voltage source.
1.
After pressing the
RANGE
button, press
ENTER
. The current range setting
will now appear. If, for example, the conditioner is set for a full-scale input of
±0.5 volts, the display will show
2.
To change the input range, you can cycle forwards or backwards through the
allowed settings (.5, 1, 2, 5, 10, and 20 volts) by pressing the top or bottom
segment, respectively, of any digit display (even if it is blank).
3.10
3
S
ETUP
: I
NSTRUMENT
C
ONFIGURATION
3.a C
ONFIGURING
T
HROUGH THE
F
RONT
P
ANEL
Содержание 3500 Series
Страница 1: ...3500 INSTRUMENT SERIES SB 3 INSTRUCTION MANUAL...
Страница 38: ...2 20 This page intentionally blank...
Страница 84: ...4 22 This page intentionally blank...
Страница 92: ...5 8 This page intentionally blank...
Страница 112: ...B 12 This page intentionally blank...
Страница 114: ...C 2 This page intentionally blank...
Страница 121: ...F 3 This page intentionally blank...