Sensing Technology
Two-electrode amperometric technology with gas permeable membranes
(replaceable)
Measuring range
0-10mg/L or 0-20mg/L
Accuracy & Resolution
+/- 0.1 mg/L
Repeatability
+/- 0.05 mg/L (25 oC)
Response Time
T90 < 90s (25 oC)
Working Condition
0 .. 50 oC with pH = 5 .. 9
Working Pressure
0 .. 15 psig
Flow range
30 .. 60 L/h or 0.6 .. 1.3 cm/s
Output
4-20mA
Power supply
6..12VDC, avg. < 200mA
Conditioning
New, first start-up: at least 4hours Restart-up: < 30 min.
Calibration
One Point - Manual with DPD
Interferences
ClO2, Ozone, Bromine Cyanuric Acid compatible
Process connection
1/2" NPT
Wetted parts
PVC, Silicone, ABS, PES
Silver-Silver Halide/Platinum
Sensor Cable
6m with BNC connector
Rating
IP68
Sensor Dimension
D21.3 x 219 (mm)
Sensor net weight
< 200 grams
Warranty
1 year for probe, membrane for 03 months
The Chlorine Electrode is designed to operate with 9-24 VDC power. Connecting the electrode to any power outside the
normal range may result in electrode damage and void the warranty.
In addition to complying with the instructions included in this manual, be sure to follow the applicable
analyzer/controller/meter instructions regarding electrode wiring.
The membrane cap needs to be filled with Internal Filled Gel(IFG) before using the electrode.
Safety Data Sheets (SDS) for information on safe handling of chemicals for operating the electrode should be made
available to all personnel involved in installation, testing, and ongoing use, as needed.
2. Specification
3. Installation
IMPORTANT NOTES: