- 7 OF  13 - 

On  the  back  side  of  the  top  of  these  light  therapy  units  there  is  an  opening  in  the  middle  below  the  air  vents   

 (see illustration). 
 At this opening the units can be fixed on the wall with a screw. Depending   
 on the type of wall, an appropriate screw must be used, may be with pegs.  
 E.g. a round head wood screw which is screwed into an expansion anchor.  
 The screw and the anchor are not included in the scope of delivery. 
 
 The screw head must not exceed a max. height of 6 mm and a max. diam  
 eter of 8 mm. The distance from the wall to the screw head can be max. 8   
 mm. The screw shaft may have a maximum diameter of 4 mm. The load on 

the screw and the condition of the wall, to which the unit has to be mounted must have a capacity of at least 3 kg. 
 
The 

light therapy unit CleanLite CL 70

 has on the back a fixture to swing out. This has to be removed before the fastening. 

Therefore 4 screws must be removed completely. Afterwards the screws have to be screwed in for safekeeping. 
The base of the 

light therapy unit CleanLite CL 110

 must be unscrewed before the fastening. Therefore the right and the 

left thumb screw must be removed completely from the foot. So the foot can be removed completely from the upper part of 
the unit. Afterwards the srews have to be screwed in for safekeeping. 

Frequency and time of application, side effects 

The treatment in the practice should be carried out 3 – 5 times per week for approx. 20 minutes by the BL 110 and approx. 
30 minutes by the BL 70 until the skin appearance has improved. Subsequently, as required, as soon as the symptoms 
appear again. At home, the treatment can take place daily. 
By using phototherapy the following side effects have been observed: Headaches, nausea, irritability and burning of the 
eyes can occur rarely for a short period. An absolute contraindication that means circumstances that do not allow the use 
of light therapy units is not known so far.  
Please take care in case of all types of eye diseases and the eye specialist should be consulted before. In case of patients 
being under treatment with lithium or medicine for example Fluoxetine, Imipramine et al., which increase the sensibility of 
the eyes on light a supervision of the light therapy by an eye specialist could be considered. 
These possible side effects / contraindications listed above can be avoided, if a protection goggle is worn during the appli-
cation. If there is as side effect an itchy skin rash further treatments should be made only in consultation with the derma-
tologist. 

Cleaning, maintenance and recycling 

The instruments do not need any special maintenance. The manufacturer does not require such preventive inspections 
by other persons. The device requires no calibration. The device is not repairable and contains no user serviceable parts. 
No modification of this equipment is allowed. If the ME equipment is in use, servicing and maintenance is not allowed. The 
mains have to be disconnected, before opening the devices. To clean the outer casing parts of the devices, we recommend 
a mild cleaning agent (as used for glass). 

 
Packaging: DAVITA

®

 takes part in the system „Grüner Punkt“ („Green Point“). All packaging parts are made out of 

recyclable materials and must be recycled in accordance of the guideless concerning each material (eco-friendly). 
The casing and the front of the cover are made of plastic material. Fluorescent lamps as well as instruments 

marked  with  the  symbol  of  a  crossed  out  dustbin  may  not  be  disposed  via  the  regular  garbage. 
 
 The electrical and electronic components must be recycled separately. Please get in touch with your local authority to find 
the right way to of disposal. If electrical appliance is disposed of in landfills or dumps, hazardous substances can leak into 
the groundwater and get into the food chain, damaging your health and well-being. For the return and disposal of batteries 
the battery law is valid. National wide there are re-sampling points. The batteries must be thrown into the green collection 
boxes. 
 
 
 
 
 
 

Содержание CleanLite CL 70

Страница 1: ...ukte GmbH Co KG Im Hammereisen 29 A D 47559 Kranenburg Lichttherapieger t Lighttherapyunit AppareildeLuminoth rapie CleanLiteCL70 CL110 Gebrauchsanweisung Usermanual Mode d emploi Deutsch 2 English 6...

Страница 2: ...den Einsatz in der N he von entflammbaren an sthetischen Mischungen mit Luft oder mit Sauerstoff oder Lachgas Das Ger t sollte frei stehen und nicht direkt der Sonne ausgesetzt werden Vor Anwendung d...

Страница 3: ...Anwendung der Lichttherapie verbieten bestehen bisher nicht Bei allen Formen von Augenerkrankungen ist Vorsicht geboten und es sollte der Augenarzt vorher aufgesucht werden Bei Patienten mit Lithiumbe...

Страница 4: ...esicherung 1 x Ger tesicherung 250 V 3A 5 x 20 Die Sicherung befindet sich hinter dem Frontdeckel Elektronisches Vorschaltger t Nach flackerfreier Z ndung sofort angenehmes flimmerfreies Licht ohne St...

Страница 5: ...sorgung Dieses Produkt nicht mit dem normalen Hausm ll entsorgen Es ist erforderlich dieses Altger t separat zu entsorgen EMC Decla ration Link http www davita de download center html Achtung Spannung...

Страница 6: ...e is not suitable for use in the presence of flammable anesthetic mixtures with air or with oxygen or nitrous oxide The instruments must be standing free and should not be exposed to sun Before using...

Страница 7: ...not known so far Please take care in case of all types of eye diseases and the eye specialist should be consulted before In case of patients being under treatment with lithium or medicine for example...

Страница 8: ...20 kWh 0 02 per hour 0 02 per hour Fuse 1 x fuse 250 V 3A 5 x 20 The fuse sits behind the front housing Electric ballast units Instant starting flicerfree light no stroboscopic effects Buzz free opera...

Страница 9: ...o not dispose this product as unsorted municipal waste Collection of such waste separately for special treatment is necessary EMC Decla ration Link http www davita de download center html Warning dang...

Страница 10: ...e plane et branchez la fiche la prise L appareil de luminoth rapie BeautyLite BL 70 est muni d un dispositif d pliant situ au dos permettant de se tenir debout en toute s curit Prenez place env 30 cm...

Страница 11: ...apie Les effets secondaires contre indications ci dessus peuvent tre vit s en utilisant des lunettes de protection port es lors de l utilisation des appareils de luminoth rapie En cas d apparitions d...

Страница 12: ...le 250 V 3 15 A 5 x 20 Le fusible se situe l arri re du couvercle avant Ballast lectronique D s allumage sans vascillement pour profiter imm diatement d une lu mi re agr able sans scintillement sans e...

Страница 13: ...ant SN Num ro de s rie Elimination des d chets Ne pas jeter ce produitsavec les d chets m nagers usuels Il est n cessaire de jeter l appareil us s par ment Decla ration EMC Lien http www davita de dow...

Отзывы: