APPLICATION NOTE

INFUTEST

Series C & Series D

AN-17.2004: Page 4 of 4

1-800-667-6557

www.datrend.com

© 

 Datrend Systems Inc.

Infutest 2000 Series C Remote Control Commands

Command description

CHAN A

Command

CHAN B

Command

Data transmitted from Infutest

to medTester

Start a Single Rate Test

RTA1

RTB1

*

Start a Dual Rate Test

RTA2

RTB2

*

Start a PCA Pump Test

RTA3

RTB3

*

Start an Occlusion Pressure
Test

RTA4

RTB4

*

Start AutoSequence number X

RSAX, X=1, 2...9

RSBX, X=1, 2...9

*

Stop a Test

STA

STB

*

Download Single Rate Test
Summary Data

GSA1

GSB1

hh:mm:ss VVVVV ml AAAAA ml/h

where   hh:mm:ss = infusion time

   VVVVV = infused volume
   AAAAA = average infusion rate

Download Dual Rate Test
Summary Data

GSA1 for RATE#1

GSA2 for RATE

#2

GSB1 for RATE#1

GSB2 for RATE

#2

hh:mm:ss VVVVV ml AAAAA ml/h

where:
hh:mm:ss = infusion time
VVVVV = infused volume
AAAAA = average infusion rate

Download PCA Pump Test
Summary Data

GSA3

GSB3

VVVVV ml AAAAA ml/h mm:ss

where:
VVVVV = average bolus volume
 AAAAA = avg bolus delivery rate
 mm:ss = average lockout time

Download Occlusion Pressure
Test
Summary Data

GSA4

GSB4

<type> PP.P psi pppp mmHg at mm:ss

where:
<type> = MAX, NRS or OVP
PP.P = occlusion pressure, PSI
pppp = occlusion pressure, mmHg
mm:ss = time of occl. pressure

Identify instrument connected to
COM2

IDENT

IDENT

INFUTEST 2000,VER X.X,N

where:
X.X = Infutest software version
N = # of channels (1, 2, or 4)

Notes:

1.

To  control  RSM  channels  C  and  D, replace "A" and "B" characters in commands above with "C" and "D"
respectively.

2.

For Run Sequence ("RS") commands, ensure that Infutest AutoSequence parameter 

"OUTPUT TO

" is set

to "Printer". Alternatively, AutoSequence 

AUTO-OUTPUT

 must be set to "Off".

3.

Do NOT use Infutest "DL" or "GR" commands with medTester.

Содержание medTester 5000B

Страница 1: ...d check items which begin with M followed by an Infutest remote control command A check item beginning with M tells medTester to get data from COM2 instead of from the medTester keyboard In executing a Checklist with M items the medTester will transmit whatever characters follow M to the Infutest via COM2 and then will wait for data to be returned via COM2 Depending on the command which follows th...

Страница 2: ...this process adding the remaining INFUTEST and DELAY commands to Sentinel s Master List Once you have finished entering all the new Check Items use the wy keys to scroll through the Master List and tag your new Check Items in the correct order by pressing the k bar Press u to copy the tagged Check Items into your new Checklist INFUTEST SR1 Exit the PROCEDURE ITEMS MAINTENANCE screen by pressing th...

Страница 3: ...negative medTester s RTS signal will prevent Infutest from downloading data To overcome this problem on the medTester you can modify your Datrend SSC 1 cable by opening the DB 25 shells and cutting the wires connected to pins 4 and 5 Solder a jumper lead so that pin 4 of the SSC 1 is shorted to pin 5 Perform this modification at both ends of the SSC 1 This will override Infutest s RTS CTS signalli...

Страница 4: ... 2 hh mm ss VVVVV ml AAAAA ml h where hh mm ss infusion time VVVVV infused volume AAAAA average infusion rate Download PCA Pump Test Summary Data GSA3 GSB3 VVVVV ml AAAAA ml h mm ss where VVVVV average bolus volume AAAAA avg bolus delivery rate mm ss average lockout time Download Occlusion Pressure Test Summary Data GSA4 GSB4 type PP P psi pppp mmHg at mm ss where type MAX NRS or OVP PP P occlusio...

Отзывы: