Tec 7 Vaporizer
4-10
1175-0013-000
Draining procedure
with Quik-Fil
w
w
w
w
CAUTION
Do not allow the container to become completely full during draining
procedures.
The vaporizer must only be drained into a properly marked container.
Step 1
Remove the cap from the vaporizer
filler.
Figure 4-13 • Unscrewing the drain plug
Figure 4-14 • Replacing the filler cap
Step 2
Place the empty container under the
drain nozzle as shown in Fig. 4-13.
Screw bottle onto drain spout.
Step 3
Unscrew the drain plug with the tool
attached to the filler cap. Drain the
vaporizer until empty.
Step 4
After draining, tighten the drain plug to
help minimize leaks.
Step 5
Replace and tighten the filler cap to
minimize leaks.
AB80026
AB80013