Page 12
Tec 5 Vaporizer
August 1999
O & M Manual Part No. 1105-0100-000
2
1
3
5
4
Fig. 11 Draining A Vaporizer Which Incorporates A Quik-Fil™ Filler
4.6.7
Draining A Vaporizer Which Incorporates A Quik-Fil™ Filler
w
Caution: Do not allow a bottle to become completely full during draining procedures.
The vaporizer must only be drained into a properly marked container, as follows:
1.
Remove the cap from the vaporizer
filler.
2.
Remove the protective cap from an
empty bottle. Insert the bottle nozzle
into the drain funnel. Rotate the
bottle to align the index slots in the
drain funnel (1) with the bottle filler
nozzle keys (2) and screw the drain
funnel onto the empty bottle as
illustrated on Fig. 11.
3.
Fully insert the drain funnel into the
keyed drain slot (3) in the bottom of
the vaporizer filler block.
4.
Unscrew the drain plug (4) with the
key (5). Continue to drain the
vaporizer until empty.
5.
After draining is complete, tighten
the drain plug to minimise the
possibility of leaks and withdraw
the drain funnel.
6.
Replace the filler cap and tighten it
to minimise the possibility of leaks.
7.
Unscrew the drain funnel from the
bottle and refit the protective cap.
Note: If further drain funnels are required
they should be obtained from the local
Abbott Laboratories representative.
AA.13.058
AA.13.057
AA.13.059