Tec 5 Vaporizer
Page 9
O & M Manual Part No. 1105-0100-000
August 1999
4.6.4
Filling A Vaporizer Which Incorporates A Keyed Filler
w
Caution: Always hold the bottle below the level of the end of the bottle adaptor and also below the
level of the filler port until the adaptor is clamped into position.
w
Caution: If it is suspected that there is liquid in the adaptor air tube, remove the bottle adaptor from
the bottle and carefully shake the adaptor two or three times to clear the tube.
The filling system consists of three elements, the bottle collar, the bottle adaptor and the filling and
draining unit fitted to the vaporizer. The vaporizer must only be filled using the correct agent specific
filling system, as follows:
1.
Screw the agent specific bottle adaptor firmly on to the bottle.
2.
Ensuring that the bottle (1) is held below the level of the end of the bottle adaptor and below the
level of the filler port, insert the end of the bottle adaptor into the filler port (2) as illustrated on
Fig. 8.
3.
Tighten the locking clamp by pulling the locking clamp lever forward and downward (3) until the
bottle adaptor is clamped.
w
Caution: When it is set correctly, the clamp secures the bottle adaptor without using the full travel
of the locking lever. The lever must not be forced to its lowest limit of travel.
4.
Raise the bottle above the level of the filler port (4) and then open the filler port valve by pulling the
port valve lever forward (5) to its full extent.
3
2
6
4
5
1
021A T5 OM
Fig. 8 Filling A Vaporizer Which Incorporates A Keyed Filler
AGENT SPECIFIC
BOTTLE COLLAR
BOTTLE
LOCKING
CLAMP
LEVER
PORT
VALVE
LEVER
AGENT SPECIFIC
BOTTLE COLLAR
LOCKING
CLAMP
LEVER
AGENT SPECIFIC
BOTTLE ADAPTOR
AA.13.046
AA.13.045