S/5 EEG Module, M-EEG and S/5 Headbox, N-EEG
Document No. 800 1011-1
19
5.1
EEG & EP service menu
EEG & EP
Check Electr.
Headbox measures impedance of electrodes. Works same way as from EEG menu or
headbox/module button.
Start AEP
AEP measurement is started with current AEP settings.
Module
Service data is changed to module view (some of the data in this view is though from headbox).
Measurement
shows whether the EEG measurement is ON or OFF. Measurement should start by
itself when the leads have been on about 15 seconds.
Montage
shows the active montage in monitor.
HB Montage
shows the montage attached to headbox. 0 means no montage, 1 basic, 2 general,
3 AEP, 4 Mont4 etc. up to 8, which means Mont8.
Montage type
shows whether the montage is bipolar or referenced.
EP Start/Stop
is highlighted when the EP Start/Stop button is pressed in headbox or module.
Imp. Check
is highlighted when the Imp. Check button is pressed in headbox or module.
Imp. Meas.
shows 1 if the impedance measurement is on in the module.
Headbox
Off
shows 1 if the headbox cable is not connected to module.
EP Meas.
shows 1 if the evoked potential measurement is active in module.
Sending EP
shows 1 if the module has acquired 100/200 new epochs and is sending them to
monitor.
Par. Error
shows 1 if the evoked potential parameters active in module are conflicting with each
other.
Reserved 1
is reserved for future use.
Reserved 2
is reserved for future use.