4.2
Troubleshooting flowchart
Possibly faulty
COP module
Take all modules out
of the Frame
Insert M-COP and turn
power on
Does fault
still appear?
Fault not in COP
module
No
Module-ID
on screen?
Does another
module work
in the same slot?
Yes
No
No
Check front panel key
functions in Service
menu
Do they work?
Yes
Check front panel unit
and COP board to
find culprit. Replace if
necessary
Return module to factory
for repair
Yes
No
Yes
Enter the Service menu
Figure 7
Cardiac Output Module troubleshooting flowchart