Datex-Ohmeda S/5 monitors
Document No. 800 1012-1
16
•
the front cover and the front panel sticker are intact
•
all connectors are intact and are attached properly
•
the module box, latch and spring locking pin are intact
?
•
Reattach the module box and check that the latch moves properly.
•
Turn the monitor on and wait until the monitoring screen appears.
•
Configure the monitor screen so that all the required parameters are shown, for example :
Monitor Setup
- Screen 1 Setup - Waveform Fields - Field 4 - P4
Digit Fields - Field 4 - SvO
2
•
Preset the C.O., SvO
2
and InvBP measurement settings:
Others
- C.O. View - C.O. Setup - Scale - 1.0 °C
Injectate Volume - 10 ml
Measurement Mode
- SET
SvO
2
- Update Hgb - 115 g/l
Invasive Pressures
- P4 Setup - Label - PA
3.
Plug in the module. Check that it goes in smoothly and locks up properly.
?
4.
Check that the module is recognized by entering the C.O. View menu:
Others
- C.O. View
Check that the message “No Catheter” is shown in the middle of the menu and the message
“No cable” in the digit field for SvO
2
, if it is an M-COPSv module.
?
5.
Enter the service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
- Service
(password 26-23-8)
Take down the information regarding COP software by selecting SCROLL VERS and turning
the ComWheel.
?
6.
Enter the COP module service menu:
Parameters - COP
Check that the “Timeouts”, “Bad checksums” and “Bad c-s by mod” values are not
increasing faster than by 50 per second. Check that the module memories have passed the
internal memory test, i.e. “RAM”, “ROM” and “EEPROM” all show OK.