![Datex-Ohmeda F-LM1 Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/datex-ohmeda/f-lm1/f-lm1_technical-reference-manual_513835050.webp)
Datex-Ohmeda S/5 Light Monitor
Document No.
M1051339
42
the displayed pressure value in order to determine the real pressure.
40.
Check the watchdog timer activation pressure.
Select
Pneumatics
from the
NIBP
service menu.
Keep the pressure manometer connected to the cuff connector. Pump up the pressure very
slowly and note the value on the manometer when you hear a signal from the loudspeaker.
The pressure at where the watchdog timer should activate with an audible signal is
13 mmHg (11…15 mmHg).
Adjust the limit with the trimmer on the NIBP board, if necessary.
41.
Check the watchdog timer.
Select
Watchdog
from the
NIBP
service menu.
Check the watchdog timer in the adult mode. Activate the timer by highlighting
Test Adult
and then pressing the ComWheel. Check that the time beside the text “Watchdog Interval”
starts to run. Wait until you hear a signal from the loudspeaker and then check the time
again. The time from the adult test should fall within 120-140 seconds.
Check the watchdog timer also in the infant mode by first selecting
Test Infant
from the
menu. The time from the infant test should fall within 60-70 seconds.
42.
Check the safety valve.
Select
Safety Valve
from the
NIBP
service menu.
Keep the pressure manometer connected to the cuff connector.
NOTE: Make sure your pressure manometer can be used to measure pressures over 300
mmHg. If such a pressure manometer is not available, perform the check with an adult cuff
that is connected around some round object, for example a calibration gas bottle.
Highlight
Start Test
. Start the adult safety valve test by pressing the ComWheel. Wait until
the pump stops and the pressure is deflated. Check the pressure values “Max press” and “2
s after stop” for both transducers. All the values should be within 290 - 330 mmHg.
Highlight
Adult
. Press the ComWheel and check that the text changes now to INFANT. Select
Start Test
and wait until the pump stops and the pressure values on the screen have been
updated. Check that the values “Max press” and “2 s after stop” are all now within
154...165 mmHg.
Return to the normal monitoring mode by pressing
Normal Screen
.
43.
Connect an adult NIBP cuff to the cuff connector and disconnect one of its hoses.
Start NIBP measurement by pressing the key
NIBP
, highlight
Start Manual
and press the
ComWheel and check that the message “Cuff loose” appears on the screen within 30
seconds.
Reconnect the hose and then bend it with your fingers. Restart the measurement and check
Содержание F-LM1
Страница 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Страница 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Страница 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Страница 78: ......
Страница 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Страница 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Страница 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Страница 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Страница 94: ......
Страница 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Страница 111: ...Service menu Document No M1051339 15 4 SERVICE MENU There is no service menu available for the CPU board ...
Страница 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Страница 114: ......
Страница 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Страница 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Страница 136: ......
Страница 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Страница 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Страница 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Страница 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Страница 206: ......
Страница 208: ......
Страница 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 224: ......
Страница 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 232: ......
Страница 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Страница 244: ......
Страница 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Страница 253: ...Service menu Document No M1051340 7 6 SERVICE MENU There is no service menu available for the power adapter ...
Страница 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Страница 256: ......
Страница 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Страница 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Страница 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Страница 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Страница 286: ......
Страница 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Страница 304: ......
Страница 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Страница 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Страница 325: ...Spare parts Document No M1051358 19 ...
Страница 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Страница 330: ......
Страница 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Страница 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Страница 342: ......
Страница 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Страница 350: ...Datex Ohmeda S 5 Light Monitor Data card Figure 7 Only Data card no menu card allowed Document No 896552 2 6 ...
Страница 352: ......
Страница 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Страница 360: ......
Страница 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Страница 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Страница 368: ......