Datex-Ohmeda S/5 Light Monitor
6
Document No. 8000060-1
2.1.4 Temperature
The temperature is measured by a probe whose resistance varies when the temperature changes,
called NTC (Negative Temperature Coefficient) resistor.
The resistance can be measured by two complementary methods:
1.
Applying a constant voltage across the resistor and measuring the current that flows
through it.
2.
Applying a constant current through the resistor and measuring the voltage that is
generated across it.
In Datex-Ohmeda solution the two methods are combined in a form of a voltage divider. The NTC-
resistor is connected in series with a normal resistor and a constant voltage is applied across them.
The temperature dependent voltage can be detected at the junction of the resistors, thus producing
the temperature signal from the patient. The signal is amplified by analog amplifiers and further
processed by digital electronics.
2.1.5 Invasive blood pressure
To measure invasive blood pressure, a catheter is inserted into an artery or vein. The invasive
pressure setup, consisting of connecting tubing, pressure transducer, an intravenous bag of normal
saline all connected together by stopcocks, is attached to the catheter. The transducer is placed
level with the heart, and electrically zeroed.
The transducer is a piezo-resistive device that converts the pressure signal to a voltage. The monitor
interprets the voltage signal so that pressure data and pressure waveforms can be displayed.
2.1.6 Respiration
Impedance respiration is measured across the thorax between three ECG electrodes. The signal of
the respiration is made by supplying current between two electrodes and by measuring the
differential current from the third electrode. The input current is 200
µ
A (31 kHz). The impedance
measured is the impedance change caused by breathing. When the patient is breathing or is
ventilated, the volume of the air in the lungs changes, resulting the impedance between the
electrodes. From these impedance changes, respiration rate is calculated, and respiration
waveform is displayed on the screen.
Содержание F-LM1
Страница 4: ...Datex Ohmeda S 5 Light Monitor Document No M1037863 2 ...
Страница 8: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 iv ...
Страница 68: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 60 ...
Страница 78: ......
Страница 80: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 ii ...
Страница 88: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 8 ...
Страница 89: ...Appendix A Planned maintenance form APPENDIX A 9 Document No 8001934 2 ...
Страница 90: ...Datex Ohmeda S 5 Light Monitor Document No 8001934 2 10 ...
Страница 94: ......
Страница 96: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 ii ...
Страница 111: ...Service menu Document No M1051339 15 4 SERVICE MENU There is no service menu available for the CPU board ...
Страница 112: ...Datex Ohmeda S 5 Light Monitor Document No M1051339 16 ...
Страница 114: ......
Страница 116: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 128: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 12 ...
Страница 134: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 18 ...
Страница 136: ......
Страница 140: ...Datex Ohmeda S 5 Light Monitor 2 Document No 8000060 1 ...
Страница 200: ...Datex Ohmeda S 5 Light Monitor 62 Document No 8000060 1 ...
Страница 201: ...APPENDIX A Check form 63 Document No 8000060 1 APPENDIX A ...
Страница 202: ...Datex Ohmeda S 5 Light Monitor 64 Document No 8000060 1 ...
Страница 206: ......
Страница 208: ......
Страница 210: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 224: ......
Страница 226: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 232: ......
Страница 234: ...Datex Ohmeda S 5 Light Monitor Document No 890201 5 ii ...
Страница 244: ......
Страница 246: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 ii ...
Страница 253: ...Service menu Document No M1051340 7 6 SERVICE MENU There is no service menu available for the power adapter ...
Страница 254: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No M1051340 8 ...
Страница 256: ......
Страница 258: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 890200 5 ii ...
Страница 274: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 ii ...
Страница 283: ...External Battery Charger N LCHGR 7 2 Exploded view Document No 896192 1 9 ...
Страница 284: ...Datex Ohmeda S 5 Light Monitor and Light Solo Document No 896192 1 10 ...
Страница 286: ......
Страница 288: ...Datex Ohmeda S 5 Light Monitor Document No 8000060 1 ii ...
Страница 302: ...Datex Ohmeda Light Monitor Document No 8000060 1 14 ...
Страница 304: ......
Страница 306: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank Document No M1051358 ii ...
Страница 316: ...Datex Ohmeda S 5 Light Monitor 65 66 64 Document No M1051358 10 ...
Страница 325: ...Spare parts Document No M1051358 19 ...
Страница 328: ...Datex Ohmeda S 5 Light Monitor This page intentionally left blank ...
Страница 330: ......
Страница 332: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 ii ...
Страница 340: ...Datex Ohmeda S 5 Light Monitor Document No 896551 2 8 ...
Страница 342: ......
Страница 344: ...Datex Ohmeda S 5 Light Monitor Document No 896552 2 ii ...
Страница 350: ...Datex Ohmeda S 5 Light Monitor Data card Figure 7 Only Data card no menu card allowed Document No 896552 2 6 ...
Страница 352: ......
Страница 354: ...Datex Ohmeda S 5 Light Monitor Document No 896553 2 ii ...
Страница 360: ......
Страница 362: ...Datex Ohmeda S 5 Light Monitor Document No 896554 2 ii ...
Страница 367: ...Method of Replacing the Lamp Unit LCD Backlight ...
Страница 368: ......