Aespire View
4-2
M1132382
Every day before your first patient
Check that necessary emergency equipment is available and in good
condition.
Check that the equipment is not damaged and that components are
correctly attached.
Check that pipeline gas supplies are connected and cylinders are
installed and adequately filled.
Check the system suction connections.
Check the vaporizer installation:
•
The top of each vaporizer is horizontal (not on crooked).
•
Each vaporizer is locked and cannot be removed.
•
The alarms and indicators operate correctly (Tec 6 series
vaporizers).
•
More than one vaporizer cannot be turned on at the same time.
•
All vaporizers are full.
Check that the breathing circuit is correctly connected, not damaged,
and the breathing system contains sufficient absorbent.
Turn the System switch to On.
Connect the scavenging and verify proper operation.
Do the Pipeline test and Cylinders test.
Do the Flow control tests.
Do the Vaporizer back pressure tests.
Do a Low-pressure leak test.
Do the Alarm tests.
Do the Breathing system tests.
Set the appropriate controls and alarm limits for the case.
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...