10.4. Messages of critical errors
Such errors are caused normally by structural injuries or
damage. Repair is done by the service of Datcon. In the
case of an error it is recommended to make a note of the
displayed error message, as well as of the phenomenon
seen and communicate that information to the Datcon
service personnel.
Default Factory Settings
(
S
ervice:
D
efault
F
actory
S
ettings)
The instrument needs to be re-calibrated, and the operating
parameters have also got damaged.
Repairing
: by service
The last saving was not successful
(
E
rror:
L
ast
S
av
e
)
Saving of the parameters modified last time was
unsuccessful. The error was caused most probably by an
interruption of the current loop. The status prior to the
modification was preserved.
Repairing
: Enter the menu, and check the parameters.
Change the incorrect values to the desired values, and
then exit the menu in the regular way. Pay attention to
ensure that the loop current should not get interrupted
during the operation.
46
20180905-V120180905-V1
Содержание DT9102 B
Страница 1: ...DT9102 B Intrinsically Safe Digital Clock Operating Instructions...
Страница 11: ...DT9102 B 4 2 Main dimensions of the instrument 20180905 V1 11...
Страница 44: ...DT9102 B 10 2 Application example 44 20180905 V120180905...
Страница 53: ...DT9102 B 10 8 ATEX Certification 20180905 V1 53...
Страница 54: ...DT9102 B 54 20180905 V120180905...
Страница 55: ...DT9102 B 20180905 V1 55...
Страница 56: ...DT9102 B 56 20180905 V120180905...
Страница 57: ...DT9102 B 20180905 V1 57...
Страница 58: ...DT9102 B 58 20180905 V120180905...
Страница 59: ...DT9102 B 20180905 V1 59...
Страница 60: ......