Repair and pcb check 

12.3

 

12.3.1 Repair 

A procedure is given to determine whether the electronics is defective in case of faults. 
If the cause is outside the electronics repair is only possible at Datawell generally. 
No procedure is given to determine the fault on the print. 
If the print is defective, the pcb should be replaced as a whole. 
All indicated measurements are with regard to "common" unless otherwise stated. 
Connected to common are: point 7 of junction box, brown leads to octal plug and all tabs 
marked "

" on pcb.

 

 
Excessive zero offset (pitch-roll), (for calibration see 10). 
Connect the pitch output to x input of an xy recorder. 
Connect the rolloutput to y input of an xy recorder. 
Sensitivity 100 mV/cm. 
Lift the Hippy gently from the ground with two men and rotate it 360° in approx. 10 seconds 
around vertical axis and replace on ground again. 
Let the xy recorder write from now on for approx. 20 minutes. 
You will observe that the xy recorder writes a nearby circle-like figure. 
The radius of this circle is a measure for the platform offset. 
 
For example:  If radius of circle 

:  2 cm 

 

Sensitivity of xy recorder  :  100 mV/cm 

 

Sensitivity of Piro 

:  174 mV/° 

 

Platform offset 

:  2x100/174 = 1.15° 

 
If no xy recorder is available, write pitch and roll output on x, t recorder for 20 minutes after 
rotation or measure and note pitch and roll output every 30 seconds. 
Pitch and roll can be plotted on graphic paper now. 
If the excessive zero offset is caused mainly by a platform offset, the instrument should be 
returned to Datawell for repair. 
If the zero offset is much larger than the platform offset, the cause may be in the electronics 
circuit board. 
Check the humidity (see 12.2). 
If the humidity was too large, dry out silicagel and measure again. 
If still too large check the print according to 12.3.2. 
 
Sensitivity error, (for calibration see 9.2 and 10). 
If sensitivity is much too small, check the connection supply-junction box and junction box-
indicating meter. 
Check if the loading resistance of the outputs is not smaller than 4 KOhm. 
The power drain should be approx. 150 Mw (measure at least 1 minute after energising).  
Check if octal plug under lid is connected properly and brown lead is plugged on screen and 
check connection octal plug-pcb. 
If everything is in order proceed according to 12.3.2. 

Содержание Hippy 40-MkII

Страница 1: ...March 20 2019 Datawell Heave Pitch Roll Sensor Manual Hippy 40 MkII from serial no 18181 Service Sales Voltastraat 3 1704 RP Heerhugowaard The Netherlands 31 72 534 5298 31 72 572 6406 www datawell nl...

Страница 2: ...2...

Страница 3: ...13 7 Transportation 14 8 Specifications 15 8 1 General 15 8 2 Pitch roll 16 8 3 Heave 17 8 4 Acceleration 18 9 System check and calibration heave 19 9 1 Simple System check 19 9 2 Calibration 19 10 Sy...

Страница 4: ...4 fig 1...

Страница 5: ...asurements is a gravity stabilised platform with a natural period time of 40 seconds An accelerometer is mounted on this stabilised platform The sensor can accept any voltage between 10 and 30 V DC Ou...

Страница 6: ...r for heave measurement 2 2 The heave is obtained by double integrating the acceleration Temporary offset of the platform caused by horizontal acceleration varying ship s speed leads to false accelera...

Страница 7: ...ls A and B fixed to housing and perpendicular to each other A pickup coil C mounted on the stabilised platform horizontal plane measures the vertical components of H1 and H2 fig 2 and 3 The induced vo...

Страница 8: ...of both planes Angle between both planes is Be H1 P the magnetic field vector generated parallel to the pitch axis OP1 the projection of OP on horizontal plane Component of H1 perpendicular to Vh PP1...

Страница 9: ...p sin ROR1 sin cos For scale factor see specifications 8 2 angle between instrument plane and horizontal plane angle between pitch axis and rotation axis POP1 angle between pitch axis and horizontal p...

Страница 10: ...Supply junction box terminal no 2 pitch junction box terminal no 3 roll junction box terminal no 4 Outputs heave junction box terminal no 5 acceleration junction box terminal no 6 common junction box...

Страница 11: ...xis are indicated by V formed cuttings in the rim of the bottom flange Mount the instrument with arrow on text plate on lid in the direction of the bow of the ship see fig 10 V formed cuttings in lid...

Страница 12: ...lane r in volts 10 sin 2 Output is positive if port side of ship is lifted The mast angle equal to angle between base plane and horizontal plane can be derived from both outputs according to 2 2 2 sin...

Страница 13: ...0 seconds is s 400 radian s peak value of horizontal displacement m For sudden changes in ship s speed v m sec is v 60 radian Platform orientation 6 3 The pitch roll measurement is independent of the...

Страница 14: ...r example when it is being hoisted may damage the accelerometer beyond repair The Hippy should not make more than six revolutions in two minutes Spinning may also lead to an entangled suspension of th...

Страница 15: ...Any voltage between 10 30 V DC 180 mW Maximum source impedance 50 Input capacitance 470 F Maximum permissible voltage between output battery and chassis 380 V Supply unit is protected against reverse...

Страница 16: ...an horizontal plane Eu 10 sin volts see fig 3 Bandwidth Low pass third order filter up to 1 Hz amplitude transfer within 1 linear phase shift equal to 0 06 seconds up to 5 Hz within 6 dB Scale accura...

Страница 17: ...accuracy gain error 0 5 of measured value after calibration 1 0 of measured value after 3 year Zero offset Below 30 cm 300 mV typically 3 cm Zero stability See under 8 2 Pitch roll Noise Below 1 cm p...

Страница 18: ...ange 10 m sec Output at terminal 6 is positive for upward acceleration Bandwidth Low pass third order filter up to 1 Hz linear phase shift equal to 0 06 sec up to 5 Hz within 6 dB Scale accuracy gain...

Страница 19: ...ending the Hippy from springs e g extenders that can be bought in a sports shop The low mechanical losses of the resulting mass spring resonant system allow one to maintain a constant amplitude oscill...

Страница 20: ...s to be 20 V 0 4 V The same applies for the roll output at roll angles of and 90 pitch axis vertical Zero offset Zero offset is equal to measured output if instrument is placed on horizontal plane Bas...

Страница 21: ...21 fig 5...

Страница 22: ...ition on top of the adjustment bolts A1 A2 A3 and A4 but with the bolts B unfixed yet Screw the bolts A1 and A3 down so that the instrument can rotate around the axis A2 A4 like a seesaw Press the ins...

Страница 23: ...23 fig 6...

Страница 24: ...e The fluid in which the stabilised platform is suspended however contains water which can diffuse very slowly through the wall of the sphere To prevent a 100 relative humidity and condensation two 65...

Страница 25: ...25 fig 8...

Страница 26: ...offset For example If radius of circle 2 cm Sensitivity of xy recorder 100 mV cm Sensitivity of Piro 174 mV Platform offset 2x100 174 1 15 If no xy recorder is available write pitch and roll output o...

Страница 27: ...responding 0 via a straight line The pitch and roll outputs should be below 0 3 50 mV if sensitivity is standard The acceleration output should be below 0 1 m sec2 0 1 V If not in order remove green a...

Страница 28: ...o the square of the frequency of the generator II Non standard sensitivity If the sensitivity is not standard there may be resistors between stand off s on pcb parallel to R157 and R158 R157a R158a R1...

Страница 29: ...resistors except R22a have to be transferred For R22a a new value has to be calculated from the value of R22a in the old pcb as follows R22a new R22a old 78 KOhm Picking a value for R22a from the E 9...

Страница 30: ...nd arrow under print pointing towards marking in the flange 2 Silicagel bags should not be in direct contact with pcb Place them in special compartments within bag see fig 6 Seal the polyethylene bag...

Страница 31: ...ly checked through perspex lid on sphere If centre of rubber membrane is pointing upwards see fig 9 B level is sufficient In case the membrane touches division plate see fig 9 A fluid has to be added...

Страница 32: ...first page manual 18001 Original 18003 Flattened flanges of can electronics changed 18033 Dimensions of housing changed Zero offset of acceleration changed into 1m sec MkII added 18121 Pcb 18 9A intr...

Страница 33: ...33 15 Dimensions and connections fig 10...

Страница 34: ......

Страница 35: ...A 1 Appendix fig 8a...

Отзывы: