10 System check and calibration pitch and roll
Scale:
Measure pitch output at pitch angle of +90° and again at pitch angle of −90°.
(Check with plummet if roll axis is vertical within 5°).
The difference of the obtained values has to be: 20 V ± 0.4 V.
The same applies for the roll output at roll angles of + and −90° (pitch axis
vertical).
Zero offset:
Zero offset is equal to measured output if instrument is placed on horizontal
plane.
Base plane and top plane are parallel within 0.1°.
So zero offset can be measured with 0.1° accuracy when horizontal top plane is
horizontal (to be checked with water level).
Do not roll instrument
Rotate via upright position
fig. 4