background image

Содержание DI-158 Series

Страница 1: ...slated into any human or computer language in any form or by any means electronic mechanical magnetic optical chemical manual or otherwise without expressed written authorization from the com pany The...

Страница 2: ......

Страница 3: ...o products that have been repaired or altered by persons other than DATAQ Instru ments employees or products that have been subjected to misuse neglect improper installation or accident DATAQ Instrume...

Страница 4: ......

Страница 5: ...ply power to your data acquisition device your PC and the device under test 4 Use a digital voltmeter to make the following measurements a Measure the voltage both the AC and DC range between the grou...

Страница 6: ......

Страница 7: ...s 6 Digital I O 7 Calibration 7 USB Interface 7 General 7 3 Installation 9 Connecting the Instrument to Your Computer 9 Installing the Device Drivers 9 Installing WinDaq Software and the Dataq Instrum...

Страница 8: ......

Страница 9: ...ment is a portable data recording module that communicates through your com puter s USB port Features include 4 fixed differential analog inputs 4 digital I O for Remote control Programmable Gain opti...

Страница 10: ...to disk while monitoring a real time display of the waveforms on screen It operates and displays waveform signals in real time at the full sample rate of the instrument being used There are two versi...

Страница 11: ...p and more in this easy to use point and click environment View the Help files by clicking on View Help in the Help menu or by pressing F1 Help All WINDAQ software utilizes context sensitive help Help...

Страница 12: ......

Страница 13: ...0V peak to peak Channel to Channel crosstalk rejection 60db Gain temperature coefficient 100ppm C Offset temperature coefficient 5 V C Output Voltage DI 148U SP String Pot Channels 2 43 typical DI 158...

Страница 14: ...ime DI 148 70 s DI 158 71 4 s Analog Outputs DI 158 Series only Number of channels 2 Resolution 12 bits Integral Nonlinearity 2 LSB Output Noise 250 Vrms Output Current 300 A Output short circuit curr...

Страница 15: ...ximum allowed 0 0 8V Calibration Calibration Cycle One year Calibration method Digital calibration with scale and offset constant per channel USB Interface Connector Type USB Maximum data transfer rat...

Страница 16: ......

Страница 17: ...addition of this new piece of hardware will be sensed by Windows and the Found New Hardware Wizard will automatically be launched anticipating the installation of a device driver for the new hardware...

Страница 18: ...elect Install Software for DI 148 DI 158 DI 710 and DI 715B instru ments and click OK 4 Select the Install Software option and click OK to continue If you wish to view the hardware documentation you m...

Страница 19: ...sers to have access to WinDaq software or just the current User Click Yes to allow all users to have access click No to allow just the current user to have access 15 After WINDAQ Software installs you...

Страница 20: ...online store or by phone For help or for questions regarding Activation and or the Software License Agreement contact DATAQ Instruments Customer Support Dataq Instruments Hardware Manager The Dataq I...

Страница 21: ...l port identification AGnd Analog Ground D5 Digital I O bit 5 D4 Digital I O bit 4 D3 Digital I O bit 3 D2 Digital I O bit 2 D1 Digital I O bit 1 Remote Event Marker D0 Digital I O bit 0 Remote Start...

Страница 22: ...I O bit 2 D1 Digital I O bit 1 Remote Event Marker D0 Digital I O bit 0 Remote Start Stop DGnd Digital Ground SP1 and SP2 string pot channels 2 channels with 3 terminals each voltage V wiper W ground...

Страница 23: ...ale range of the instrument 10V is suggested 1 Enter the Low Calibration a In the Edit menu click on Low Calibration b Enter 10 for the Input Level and 0 the string pot at position 0 for the Low Cal V...

Страница 24: ...menu click on High Calibration b Enter 10 for the Input Level and the greatest length of the displacement cable for the High Cal Value Enter the Engr Units using the proper measurement quantifier i e...

Страница 25: ...it will experience common mode voltages in excess of 30 VDC or peak AC the device s full scale range without damage Port Designations Refer to the following for connections AGnd Analog Ground DAC0 Di...

Страница 26: ...nst the signal wire and not the wire insulation Do not over tighten 3 Tug gently on the signal lead to ensure that it is firmly secured When an input signal is connected and WINDAQ Acquisition softwar...

Страница 27: ...put signal and the common mode voltage CMV If this sum exceeds the maximum rating of your device see 2 Specifications you must determine the source of the CMV and eliminate it before taking any measur...

Страница 28: ...iguration Use the following diagram for proper multi sensor process measurement connections With proper precautions process current measurements can be made with little if any risk to your data acquis...

Страница 29: ...nalog Input Diagram DI 148U DI 148U SP 10 60 4K 330pf QDORJ QSXW 7R QSXW PSHGDQFH 2 4V 46 4K SP1 SP2 CH3 CH4 General Purpose Analog In Field Side Screw Terminal 6WULQJ 3RW 63 5HVLVWDQFH Excitation 2 4...

Страница 30: ...I 148U SP and DI 158 Series Hardware Manual Connecting Input Signals 22 DI 158 Series SI SI OHFWURQLF 6ZLWFK FORVHG IRU 9 RSHUDWLRQ QSXW PSHGDQFH HLWKHU LQSXW WR 9 0 LIIHUHQWLDO 9 9 RQWURO 7R 83 QG 7R...

Страница 31: ......

Страница 32: ...Telephone 330 668 1444 Fax 330 666 5434 E mail support dataq com Vertrieb Distributed by ALTHEN GmbH Mess und Sensortechnik Frankfurter Str 150 152 Tel 49 0 6195 7006 0 http www althen de 65779 Kelkh...

Отзывы: