![Data Translation DT9834 Series Скачать руководство пользователя страница 126](http://html.mh-extra.com/html/data-translation/dt9834-series/dt9834-series_user-manual_511232126.webp)
Chapter 8
126
21.
Adjust the gain by entering values between 0 and 255 in the DAC 3 Gain edit box or by
clicking the up/down buttons until the voltmeter reads 9.375 V.
Note:
At any time, you can click
Restore Factory Settings
to reset the D/A calibration
values to their original factory settings. This process will undo any D/A calibration settings.
Once you have finished this procedure, the analog output circuitry is calibrated. To close the
DT9834 Calibration Utility, click the close box in the upper right corner of the window.