Chapter 8
108
Calibrating the D/A Subsystem
Do the following to calibrate the D/A subsystem on the board:
1.
Connect a precision voltmeter to DAC 0 (screw terminal/pin 34 for ANO0+ and 33 for
ANO0–).
2.
Set the value to the minus full-scale range (all zeros).
3.
Adjust potentiometer R5, shown in
Figure 29
, until the analog output voltage is –10 V.
4.
Set the value to the plus full-scale range (all ones).
5.
Adjust potentiometer R6, shown in
Figure 29
, until the analog output voltage is
9.995 ς
.
6.
Connect a precision voltmeter to DAC 1 (screw terminal/pin 32 for ANO1+ and 31 for
ANO1–).
7.
Set the value to the minus full-scale range (all zeros).
8.
Adjust potentiometer R13, shown in
Figure 29
, until the analog output voltage is –10 V.
9.
Set the value to the plus full-scale range (all ones).
10.
Adjust potentiometer R14, shown in
Figure 29
, until the analog output voltage is
9.995 ς
.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание DT3000 Series
Страница 5: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 11: ...Contents 10 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 21: ...Chapter 1 20 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 67: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 93: ...Chapter 6 92 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 105: ...Chapter 7 104 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 111: ...Chapter 8 110 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 131: ...Appendix A 130 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 147: ...Appendix C 146 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 155: ...Index 154 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...