Data Harvest SmartQ 3266 Скачать руководство пользователя страница 8

                                      Drop / Bubble Counter 

 

 

 

Clamp the reservoir and position in the gap 
between the receiver and emitter so that 
the reservoir is touching the emitter’s 
moulding.  

Adjust the Counter or reservoir vertically so 
the moulding ring at the bottom of the 
reservoir barrel is level with the lower edge 
of the emitter moulding.  

Adjust the reservoir laterally so the 
reservoir is just touching the moulding of 
the large hole in the alignment plate. In this 
position bubbles rising should pass through 
the light path between the receiver and 
emitter.  

Note: You may need to make fine lateral 
adjustments of the reservoir.  

Check the reservoir is vertical and parallel to 
the Counter.  

Note: Try to keep the experimental apparatus higher than 
the water level in the reagent reservoir to prevent backflow 
of water into the experimental apparatus.  

 

Pour approximately 10 cm

3

 of water into the reservoir. 

 

Test to make sure the bubbles cause the red LED 

indicator on the Counter to blink as they rise through 
the water  

 

Press the zero reset button on the Counter. Run the 

experiment (remember to turn the stopcock to an 
open position).  

Practical information  

 

The fastest speed that data can be captured is 50 Hz (20 ms). If an intersample time of 
less than 20 milliseconds is selected, then the values obtained will either default to the 
lowest reading or the set up will be rejected by the logger/software.  

 

A replacement parts pack for the Counter reservoir is available from Data Harvest, which 
contains 2 x 3-way stopcocks and 5 plastic tips - use Product No. 3271. 

 

The 

Smart Q

 Drop / Bubble Counter works by recording interruptions of a beam of 

infrared energy. The receiver can be sensitive to high levels of ambient infrared light or 
heat. Shield the Counter from bright light e.g. sunlight, which can produce a false OFF 
result. Work in an area away from direct light, or rotate through 180˚ so the external light 
source is directed towards the emitter.  

 

Any calculations for drops per cm

3

 will only be valid for the same dropper with the same 

type of solution and flow rate. 

 

If the flow rate is too fast the drops will merge to form a continuous stream. The flow rate 
should be set so that each drop passes thorough the Counter before the succeeding drop.  

 

With a titration use a very slow drop rate (e.g. with a maximum speed of 1.5 drops a 
second) so the drops can add too and mix thoroughly with the reactant, allowing the 
electrode time to respond.  

 

If you use a burette as a reservoir you will need to devise a method of stopping the flow 
from the burette without altering the drop rate. If the burette has a replaceable tip, use a 
length of plastic tubing to link the tip to the burette. A tube clamp can be used to crimp the 
plastic tubing and shut off the flow. A bung inserted into the top of a burette will also 

Reagent 
reservoir 

Drop/Bubble 
Counter 

Counter must be 
at 90º, use the 
moulding line to 
get a vertical 

Alignment 
Adapter 
 

Reservoir just   
touching the 
moulding of the 
large hole in 
the adapter 

Содержание SmartQ 3266

Страница 1: ...o uk Data Harvest Freely photocopiable for use within the purchasers establishment Drop Bubble Counter Product No 3266 Ranges Drop Bubble Count 0 10 000 23 drops cm3 Volume 0 120 cm3 24 drops cm3 Volu...

Страница 2: ...gnal that is counted by the internal counter module The red LED is an indicator which will blink as a drop or bubble passes through the infrared beam There is a small button close to the indicator tha...

Страница 3: ...ASYSENSE unit into the hooded socket on the Counter housing Connect the other end of the sensor cable to an input socket on the EASYSENSE unit The red LED will light The EASYSENSE unit will detect tha...

Страница 4: ...set the rate of flow and therefore the drop rate from the reservoir The lower stopcock is used in either an open or closed position to turn the drops on and off The line of the blue handle indicates t...

Страница 5: ...he body of the reservoir Do not over tighten Check the reservoir is vertical The drops produced will fall through and cut the infrared beam without further adjustment If the reagent reservoir and alig...

Страница 6: ...topcocks a and fill the reservoir with the solution being used 2 The first step is to adjust the flow rate Place a beaker under the stopcock to catch the drops Fully open the lower stopcock b Slowly t...

Страница 7: ...alibrated for volume but can be used to show the change of gas production over time Test your set up before beginning the experiment It is critical that the column of fluid is positioned so the bubble...

Страница 8: ...ment parts pack for the Counter reservoir is available from Data Harvest which contains 2 x 3 way stopcocks and 5 plastic tips use Product No 3271 The Smart Q Drop Bubble Counter works by recording in...

Страница 9: ...s used as a Drop Counter with the Smart Q pH sensor to record a pH vs volume graph e g sodium hydroxide with hydrochloric or ethanoic acid ammonia with hydrochloric or ethanoic acid sodium carbonate w...

Страница 10: ...o zero 8 Open the EASYSENSE program and select EasyLog from the Home page 9 If the necessary set the Drop Counter to the required range i e Drop Bubble Count or Volume in cm3 e g 27 drops cm3 10 Click...

Страница 11: ...diagram above Test to make sure bubbles rising up in the reservoir will be detected Attach the Bubble Counter to the EASYSENSE unit 2 Place 40 cm3 of active yeast solution e g 0 1 w v into the conical...

Страница 12: ...be extended to see the effect of different sugar substrates on rate as part of an investigation into respiration glycolysis If a temperature controlled water bath is accessible study the effect of hea...

Отзывы: