background image

                                      Drop / Bubble Counter 

 

 

If accuracy is not critical and you are using the reagent reservoir and tip supplied with a low 
viscosity liquid (like water) and the flow rate set to: -   

 

Fast e.g. 

10 plus

 drops per second, use the 

24

 drops/cm

3

 range.  

 

Medium e.g. between 

– 10

 drops per second, use the 

25

 drops/cm

3

 range.   

 

Slow e.g. between 

1.5 

– 5

 drops per second, use the 

26 

drops/cm

3

 range.  

 

Very Slow e.g. 

less than

 

1.5

 drops per second, use the 

27

 drops/cm

3

 range 

 

Note: When used with a pH or Conductivity Sensor, the flow rate needs to be very slow (less 
than 1.5 drops per second) to allow the Sensor time to settle to a new reading after addition of 
the titrant. 
 
To calculate the number of drops in a cm

3

 

The volume of a drop of fluid (and therefore the number of drops per cm3) depends on a 
number of factors. These include the:  

 

 

Size and shape of the dropper end  

 

Type of solution (its density, viscosity and surface tension)   

 

Flow rate of the liquid through the dropper end (the slower the dropping the smaller the drop)  

 

1.  Set up the reagent reservoir in the alignment adapter of the Counter. Close both 

stopcocks (a) and fill the reservoir with the solution being used.  

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  

 

 

 

2.  The first step is to adjust the flow rate. Place a beaker under the stopcock to catch the 

drops. Fully open the lower stopcock (b). Slowly turn the top stopcock (c) until it begins to 
produce drops and then finely adjust the drop rate. When the correct flow rate of drops is 
achieved close the lower stopcock (d) to stop the flow. Now the ‘flow rate’ is set, do not 
adjust the top stopcock 

– leave in this position. Use the lower stopcock to turn the drops 

on and off (e).  

 

Note: Do not set the flow rate too fast or the drops may form a stream and will not be counted as 
individual drops. 

 

3.  Top up the reservoir. Place the accurate measuring container e.g. volumetric flask (10 

mls or less) under the dropping tip.  

4.  Set the range of the Counter to Drop/Bubble (Count). Press the zero reset button.   

5.  Open the lower stopcock 

fully

 and count the number of drops required to fill up to the 

volume mark on the measuring container. Use the lower stopcock to turn the drops off. 

Notes: Make sure the lower stopcock is fully open or the rate of flow (and therefore the number of 
drops per cm

3

) will be affected. 

Closed 

(OFF)

 

Open 

(ON)

 

Closed

 

(OFF) 

 

Closed 

(OFF)

 

 

Closed 

(OFF)

 

 

Open 

(ON)

 

Open 

(ON)

 

Open 
slowly to 
adjust 
drop rate 

Partly 
Open 

 

Partly 
Open 

 

 

(e) 

(d) 

(c) 

(b) 

(a) 

Содержание SmartQ 3266

Страница 1: ...o uk Data Harvest Freely photocopiable for use within the purchasers establishment Drop Bubble Counter Product No 3266 Ranges Drop Bubble Count 0 10 000 23 drops cm3 Volume 0 120 cm3 24 drops cm3 Volu...

Страница 2: ...gnal that is counted by the internal counter module The red LED is an indicator which will blink as a drop or bubble passes through the infrared beam There is a small button close to the indicator tha...

Страница 3: ...ASYSENSE unit into the hooded socket on the Counter housing Connect the other end of the sensor cable to an input socket on the EASYSENSE unit The red LED will light The EASYSENSE unit will detect tha...

Страница 4: ...set the rate of flow and therefore the drop rate from the reservoir The lower stopcock is used in either an open or closed position to turn the drops on and off The line of the blue handle indicates t...

Страница 5: ...he body of the reservoir Do not over tighten Check the reservoir is vertical The drops produced will fall through and cut the infrared beam without further adjustment If the reagent reservoir and alig...

Страница 6: ...topcocks a and fill the reservoir with the solution being used 2 The first step is to adjust the flow rate Place a beaker under the stopcock to catch the drops Fully open the lower stopcock b Slowly t...

Страница 7: ...alibrated for volume but can be used to show the change of gas production over time Test your set up before beginning the experiment It is critical that the column of fluid is positioned so the bubble...

Страница 8: ...ment parts pack for the Counter reservoir is available from Data Harvest which contains 2 x 3 way stopcocks and 5 plastic tips use Product No 3271 The Smart Q Drop Bubble Counter works by recording in...

Страница 9: ...s used as a Drop Counter with the Smart Q pH sensor to record a pH vs volume graph e g sodium hydroxide with hydrochloric or ethanoic acid ammonia with hydrochloric or ethanoic acid sodium carbonate w...

Страница 10: ...o zero 8 Open the EASYSENSE program and select EasyLog from the Home page 9 If the necessary set the Drop Counter to the required range i e Drop Bubble Count or Volume in cm3 e g 27 drops cm3 10 Click...

Страница 11: ...diagram above Test to make sure bubbles rising up in the reservoir will be detected Attach the Bubble Counter to the EASYSENSE unit 2 Place 40 cm3 of active yeast solution e g 0 1 w v into the conical...

Страница 12: ...be extended to see the effect of different sugar substrates on rate as part of an investigation into respiration glycolysis If a temperature controlled water bath is accessible study the effect of hea...

Отзывы: