background image

1130 - Wireless 20V Voltage & Current Sensor

10 / 19  |  Revision: 1.0  |  DS157

Usage Information

CAUTION:

 Never connect a Current sensor directly across a battery or power supply without a

resistance component to limit the current to within the range of the sensor. Failure to limit the current
will cause permanent damage to the sensor.

·

If the Voltage sensor is not part of a complete circuit, then data collected may appear ‘noisy’. To
measure voltage accurately you need an impedance (resistance); the Voltage sensor is a high
impedance device and will pick up any electrical ‘noise’. 

·

Note: To demonstrate zero impedance, short out a Voltage sensor by connecting its black & red
plugs together.

·

A tare function is present in the software to set the zero of current and/ or voltage. 

·

The Current sensor has a very low resistance, so it will introduce as little resistance as possible to
the circuit. 

·

For some work, the d.c. output from a chemical cell will give the best results, for example,
discharge and charge constants of a capacitor resistance circuit.

·

An alternative to batteries is to use a fully isolated mains power supply with a regulated d.c. output
(smoothed and fully rectified).

·

Be aware that some power supplies are 1/2 wave rectified producing an average rather than true
d.c. The sensors will ‘pick up’ the fluctuations in voltage and current from this type of power supply.
The Voltage and Current sensor does not smooth the data, as you would see in a multimeter, and
will return the value that is present at the sample period. Without smoothing you can record at very
short intersample periods (50µs or 50,000 Hertz).

The suitability of components can be checked using Ohms law.

The Current sensor is protected to a voltage of ±13 V. The maximum working voltage of the Current
sensor before damage is ±13 V. 
For example: A 100 

Ω 3 W resistor gives good results without too much heat.

§

6 V supply with 10 

Ω resistor = 0.6 A (600 mA) current flow (Power 0.6 x 6 = 3.6

W).

§

6 V supply with 50 

Ω resistor = 0.12 A (120 mA) current flow (Power 0.12 x 6 =

0.72W).

§

6 V supply with 100 

Ω resistor = 0.06 A (60 mA) current flow (Power 0.06 x 6 =

0.36W).

·

The Current sensor is protected to a voltage of ±13 V, The maximum working voltage of the Current
sensor before damage is ±13 V.

·

The operating range of the sensor is 0 to 40 

̊C and 0 to 95% RH (non-condensing). Do not subject

to extreme heat or cold. Do not expose to direct sunlight for extended periods of time.

·

The sensor is not waterproof. It may be cleaned using a damp cloth. Do not immerse in water or
detergent. Do not place the sensor in an environment in which high humidity levels are possible as
this may result in damage or malfunction.

Содержание 1130

Страница 1: ...Copyright Data Harvest Group Ltd All Rights Reserved Freely photocopiable for use within the purchaser s establishment 1130 Wireless 20V Voltage Current Sensor Revision 1 0 DS157...

Страница 2: ...oduction 3 Pack Contents 4 Operational Overview 5 Connectivity 7 Charging the Sensor 8 Firmware Updates 9 Usage Information 10 Practical Investigations 11 Sensor Specifications 13 Limited Warranty 14...

Страница 3: ...both electric current and the potential difference across a component in low voltage a c or d c circuits or be used as a voltage or current sensor Software and firmware will allow the user to show res...

Страница 4: ...This product is supplied with the following items 1 x Wireless 20V Voltage Current Sensor 1 x USB Connecting Lead Additional Accessories To get the most from your Smart Wireless Voltage Current Senso...

Страница 5: ...between the device s internal sensor and your experiment The Status Indicators 2 The sensor features a single status indicator that changes colour and flashes See the table below for further informat...

Страница 6: ...to respond contact Product Support at Data Harvest Please provide details of o The computer platform it is being used with and the EasySense2 app s version number o A description of the problem being...

Страница 7: ...mobile phones as well as desktop or laptop computers giving students the ability to run experiments independently without being tethered to a device See the EasySense2 app user manual system requirem...

Страница 8: ...mA or more it will automatically recharge the battery LED status flashing white When connected to a computer the computer should be turned on and not in sleep or standby mode as the battery may drain...

Страница 9: ...nect your sensor to the EasySense2 app You will be given the option to decline an update Updates can be performed over USB or Bluetooth and will typically take less than one minute Updating firmware o...

Страница 10: ...ified producing an average rather than true d c The sensors will pick up the fluctuations in voltage and current from this type of power supply The Voltage and Current sensor does not smooth the data...

Страница 11: ...p Faraday s induction of current or voltage in a coil Capacitor discharge charge and energy stored Diodes a c diode rectification Specific heat capacity for a liquid or solid Calibration of a thermome...

Страница 12: ...hey work View video playlist Explore EasySense2 The core of our science platform is our EasySense2 software In these videos you will learn everything from the basics of our software to the most in dep...

Страница 13: ...000 samples per second interval 50 s Power specifications 5 V at 500 mA Connectivity Wired via USB Wireless via Bluetooth Bluetooth Specifications Bluetooth 4 2 low energy radio single mode compliant...

Страница 14: ...of purchase Postage Charges In the event of a fault developing the product must be returned in suitable packaging to Data Harvest for repair or replacement at no expense to the user other than postal...

Страница 15: ...ed as life expired and we will dispose of them for you FCC Details This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause h...

Страница 16: ...e Data Harvest support team Feature Detail Loss of Bluetooth Connectivity If the sensor loses Bluetooth connection and will not reconnect try Closing and reopening the EasySense 2 app Switching the se...

Страница 17: ...ailable to the EasySense 2 app 2 When the sensor is connected to a computer the computer should be turned on and not in sleep or standby mode or the battery may drain instead of charge 3 Data Harvest...

Страница 18: ...Eden Way Leighton Buzzard Bedfordshire LU7 4FY United Kingdom Tel 44 0 1525 373666 Fax 44 0 1525 851638 Sales email sales data harvest co uk Support email support data harvest co uk Online Communicat...

Страница 19: ...s manual is by default provided in the English United Kingdom language If an alternative translation is available it will be listed here We have for your convenience included a webpage translation fea...

Отзывы: