background image

 

Page 38 of 43 

6. Troubleshooting Guide & Spare Parts 

 

Problem 

Possible Cause 

Corrective Actions 

 

No light output 

Power Supply 

Wall Version - Check the plug top 
adapter is plugged in and switched 
on at the power socket. 

On/Off/Intensity 

Button 

Press the button on the back of the 
light head to check if the light will 
power on (if dimmed refer to poor 
light output actions). 

Fuses 

Visually check fuses then check 
continuity with meter 

Cables 

Check the visible wiring and make 
sure the DC connectors between the 
light and the extension cable are 
fully connected. For the mobile 
version, ensure that the IEC is firmly 
pushed into the PSU. 

 

Poor Light 

output 

Power Button 

Please check the variable intensity is 
not just set to its minimum. To do 
this use the intensity controls on the 
rear of the light head. 

Voltage in 

Verify that the correct mains voltage 
and correct transformer is being 
used  

Voltage out 

Verify correct secondary voltage 

 

Arm drift 

sideways 

Wall mount 

On installation the arm has most 
likely not been vertically aligned and 
levelled. Please contact the original 
fitter or a qualified electrician 

Head movement 

or drift 

Light head 

Please check the tightness of the 
thumb screw adjuster just below the 
head of the light 
 

 
 
 
 
 
 

 

Содержание X340

Страница 1: ...X340 LED Examination Light Operating Installation Manual QAM X340 0519 9...

Страница 2: ...Page 1 of 43...

Страница 3: ...ion 15 4 1 Wall mounted versions 15 4 1 1 Considerations 16 4 1 2 X340LW Placement and Installation 17 4 1 3 X340LE1 Placement and Installation 19 4 1 4 X340LE2 Placement and Installation 19 4 1 5 X34...

Страница 4: ...llation 30 4 6 7 Light head and arm fixing 33 5 Maintenance 34 5 1 Maintenance schedule 34 5 2 Safety Precautions 34 5 3 Product Cleaning Care Guidelines 35 5 4 Detaching Parts 36 5 5 Fuses and their...

Страница 5: ...e positioning of the light patch The light comprises 3 parts light head and articulated arm system power supply module and the mount of choice The rigid spring balanced arm allows the light patch to b...

Страница 6: ...pported in this Manual X340LC X340 LED Ceiling Mounted Light X340LW X340 LED Wall Mounted Light X340LR X340 LED Rail Mounted Light X340LD X340 LED Desk Mounted Light X340LDS X340 LED Weighted Desk Sta...

Страница 7: ...ing R9 88 Light Field Diameter D50 0 5m 90mm 0 75m 125mm 1 0m 175mm Light Field Diameter D10 0 5m 220mm 0 75m 255mm 1 0m 340mm Number of LED s 6 Power Consumption 14W 0 2A Average Working Life 50 000...

Страница 8: ...V Plug top transformer 1 24V Plug top transformer 1 Desk Clamp Fixed Desk Mount Desk Stand 1 Wall bracket fixings 1 Light head and rigid arm system 1 Thumb Screw for Wall Mount 1 Microfibre cloth 1 Li...

Страница 9: ...rigid arm system 1 User Manual Microfibre cloth 1 User Manual 1 2 2 Pre Start Checks Please check that the power cable is without visible damage harsh folds knots or cuts in the rubberised coating Wh...

Страница 10: ...Page 9 of 43 3 Operation 3 1 Range of Motion The full range of motion between the X340 Range varies This is an image depicting the range of motion on a standard wall mount fitting X340LW...

Страница 11: ...Page 10 of 43 X340LE4...

Страница 12: ...e 11 of 43 This is an image depicting the range of motion on a standard ceiling mount fitting X340LC Downtube Rigid Arm Light head Tension Adjustment Downtube Extension Retainer Ring Ceiling Mount Cov...

Страница 13: ...to power ON or OFF Press and hold the power button to adjust the intensity It takes approximately 5 seconds to switch between the highest and lowest settings Release the button when at a comfortable s...

Страница 14: ...itself DO NOT confuse this with the tension adjustment knob on the left side see section 5 6 Page 35 Loosening this thumb screw will allow you to pivot the light head up or down The head has a total t...

Страница 15: ...300mm see below image This extension is controlled by a thumb screw located around the back of the primary downtube Loosening the thumb screw will allow you to reposition the light to the required he...

Страница 16: ...atible versions X340LW X340LE1 X340LE2 X340LE3 X340LE4 The wall bracket needs to be true on the vertical plane PIVOTW2 supplied with X340LW PIVOTW6 supplied with X340LE4 PIVOTW5 supplied with X340LE3...

Страница 17: ...ical procedures the light will be used for If in doubt consult the end user or DARAY When installing the bracket take consideration of where you want the light head to sit when in its lowest position...

Страница 18: ...ring in mind the final position of the light head required to project its light patch where it will be needed We recommend 850 1100mm from the floor to the underside of the bracket dependent on couch...

Страница 19: ...ight on See page 12 Check arm and head for smooth and correct movement Your new DARAY light is now ready for use To gain full support of our warranty please fill in our warranty web form or contact DA...

Страница 20: ...s version includes our PIVOTW4 spreader plate wall mounting bracket Suitable for fixing to all wall types including brickwork A separate power supply is included Mark on the wall the selected location...

Страница 21: ...unting bracket Fit front bracket and rear plate to pre drilled panel Fit light spigot into bracket and pass cable through to the rear of panel Make electrical connection to transformer Fit pre drilled...

Страница 22: ...s from the wall mounting box with the hex wrench provided Feed the mains supply via the cable entry hole on the back plate Connect the mains supply to the fused terminal block found on the back plate...

Страница 23: ...il and fit the light to the rail The light can be moved along the rail by slackening the clamp slightly and not sufficiently to allow the light to become detached from the rail 4 3 Desk mounted versio...

Страница 24: ...e upstand push down with a firm pressure until the upstand is fully inserted NOTE Insertion of upstand may be very tight and may require the use of a wooden or rubber mallet to tap the upstand downwar...

Страница 25: ...e into the black power supply attached to the base upstand The light can now be tested by connecting the DC jacks between the power supply and the light and applying power Press the switch on the head...

Страница 26: ...stand terminal Connect the plug top transformer to the 5521 IN OUT connector on the battery backup unit Insert the Plug top transformer into a 13amp wall socket Turn the battery back up unit on with t...

Страница 27: ...und live circuitry should be qualified and following proper procedures Heavy or Difficult fittings Fitting the ceiling plate can be difficult and aspects of the fitting may require more than one perso...

Страница 28: ...re and secured with chemical bonding The is our primary choice for mounting our lights to the ceiling structure appropriate rawl bolts and chemical bonding should be used to secure the fitting Bolted...

Страница 29: ...Affixing the Ceiling plate The location of the light should be decided by the end user project manager or through schematical drawings Once the location has been decided it will be necessary to first...

Страница 30: ...5mm twin and earth type DARAY recommends that the power supplied uses an earthed switched fused spur protected with a 3amp fuse The above wiring is for modern wiring applications post 2000 On completi...

Страница 31: ...s of what length downtube would be required Height of Ceiling Plate from Floor Length of Downtube Required 2 4m 500 2 65m 750 2 9m 1000 3 4m 1500 3 9m 2000 The minimum length the downtube can be cut t...

Страница 32: ...vent damage to the cable Before drilling the safety bolts hole please direct spigot on the bottom the downtube toward the centre of the examination area to allow the best range of motion over the exam...

Страница 33: ...on the ceiling plate Slide the safety bolt through the clasps and the downtube being careful not to catch the wire Be sure to tighten the safety bolt and the nylon locknut It will be necessary to ens...

Страница 34: ...ther The light can now be tested by applying power and pressing the power button on the back of the head Check the arm and head for smooth and correct movement Your new DARAY light is now ready for us...

Страница 35: ...ge Remove arm clean spigot and re grease as necessary Check all electrical connections are sound and that there is no visible cable wear All fixings to be checked and be of sound construction 5 2 Safe...

Страница 36: ...our BioProtect antimicrobial coating on painted finishes This is designed to assist with infection control and is not a replacement for cleaning CAUTION Our products are NOT waterproof The equipment...

Страница 37: ...40 please contact DARAY as unauthorised work will invalidate the warranty 5 5 Fuses and their replacement For the ceiling mounted light the fuse is located on the ceiling plate on the terminal block c...

Страница 38: ...arm appears to move by itself and does not settle in the required position This can be rectified by adjusting the spring tension on the arm Next to the light head on the left side of the arm systems...

Страница 39: ...d the extension cable are fully connected For the mobile version ensure that the IEC is firmly pushed into the PSU Poor Light output Power Button Please check the variable intensity is not just set to...

Страница 40: ...Amp ceiling plate fuse CS6144 For any SPARE PARTS or TECHNICAL ISSUES phone DARAY S Service Delivery Team on 44 0 800 878 9864 Opening times 9am 5pm Monday to Thursday excluding bank holidays 9am 2 30...

Страница 41: ...red by any of the above We will always try to help but we cannot normally offer a refund For additional clarification please refer to our terms and conditions at www daray co uk terms In a small numbe...

Страница 42: ...Daray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from...

Страница 43: ...Page 42 of 43 Service Notes...

Страница 44: ...Page 43 of 43...

Страница 45: ...Page 44 of 43...

Страница 46: ...Page 45 of 43...

Страница 47: ...Page 46 of 43...

Страница 48: ...DARAY Ltd Edison House Robian Way Swadlincote Derbyshire United Kingdom DE11 9DH 44 0 800 878 9864 44 0 333 321 0973 sales daray co uk www daray co uk...

Отзывы: