1. Introduction 

 

DARAY’s BH200 LED is a state-of-the-art, low-cost patient light. The light uses the 
latest next-gen LED technology with no heat in the light beam. 
 
The BH200 LED head swivels on a flexible arm to allow accurate and stable 
positioning of the light patch. Its head and handles are coated with DARAY’s 
BioProtect™, an antimicrobial protection against bacteria such as MRSA. Wall or rail 
mountings are available.  
 
The light comprises 3 parts: light head and flexible arm system, power-supply 
module, and the mount of choice. The flexible arm allows the light patch to be 
placed accurately and stably where it is needed. The arm is covered with a PVC 
sleeve to prevent contamination and to allow easy cleaning. 
 
The plug-in power-supply module contains an electronic transformer to provide a low 
voltage to power the light from the electrical mains.  
 
The wall-mounting bracket provides a simple method of fixing the light to the wall, 
allowing the base of the arm to swivel in front of the wall; a screw prevents the arm 
from accidentally being pulled out of the bracket.  
 
The rail-mount version features a clamp to fasten the light to a medical equipment 
rail which is optionally supplied in 1m lengths. The clamp can also be used to mount 
the light to a bed or trolley rail.  
 
All versions are delivered in special-to-type packaging which identifies on the ends of 
the box the model type and serial number of the light within. 
 
This user & installation guide contains all the information you need for installation, 
operation and maintenance.

 

 

Restrictions and Liabilities 
 
Information in this document is subject to change and does not represent a 
commitment by DARAY. Changes made to the information in this document will be 
incorporated in new editions of the publication. No responsibility is assumed by 
DARAY for the use or reliability of software or equipment that is not supplied by 
DARAY. 
 
 

 

Содержание BH200

Страница 1: ...BH200 LED Patient Light Operating Installation Manual QAM BH200 0519 5...

Страница 2: ...Page 1 of 20...

Страница 3: ...On Variable Intensity 8 4 Installation 9 4 1 Wall mounted version BH200LW 9 4 1 1 Considerations 9 4 1 2 Placement and Installation 10 4 2 Rail mounted version BH200LR 12 5 Maintenance 13 5 1 Mainten...

Страница 4: ...ctrical mains The wall mounting bracket provides a simple method of fixing the light to the wall allowing the base of the arm to swivel in front of the wall a screw prevents the arm from accidentally...

Страница 5: ...www daray co uk or call 0333 321 0971 This product was designed manufactured in Great Britain by Daray Ltd Edison House Robian Way Swadlincote Derbyshire DE11 9DH 1 1 Supported Lights The Following Mo...

Страница 6: ...Index CRI 95 Colour Rendering R9 88 Light Field Diameter D50 0 5m 280mm Light Field Diameter D10 0 5m 640mm Number of LED s 3 Power Consumption 12W 0 1A Average Working Life 50 000 Hours Input Voltage...

Страница 7: ...g installation should be the following items in the specified quantities BH200LR Qty BH200LW Qty 12V Plug top transformer 1 8m 1 12V Plug top transformer 1 8m 1 Rail Clamp 1 Wall bracket fixings 1 Thu...

Страница 8: ...Page 7 of 20 3 Operation 3 1 Range of Motion This is an image depicting the range of motion on the BH200 light all versions...

Страница 9: ...and hold to adjust the intensity It takes approximately 4 seconds to switch between the highest and lowest settings Release the button when at a comfortable setting for the user Press once to power OF...

Страница 10: ...to the underside of the bracket dependent on couch height It may be necessary to adjust this height if there are specific medical procedures the light will be used for If in doubt consult the end use...

Страница 11: ...in mind the final position of the light head required to project its light patch where it will be needed We recommend 850 1100mm from the floor to the underside of the bracket dependent on couch heig...

Страница 12: ...light on See page 8 Check arm and head for smooth and correct movement Your new DARAY light is now ready for use To gain full support of our warranty please fill in our warranty web form or contact DA...

Страница 13: ...mark the positions for the 2 screw holes Fix the rail to the wall using appropriate plugs as necessary and the 2 screws provided NOTE Slide the screw head cap retainers the correct way up onto the scr...

Страница 14: ...and that there is no visible cable wear All fixings to be checked and be of sound construction 5 2 Safety Precautions For all cleaning work power off the equipment and where possible unplug the mains...

Страница 15: ...e our BioProtect antimicrobial coating on painted finishes This is designed to assist with infection control and is not a replacement for cleaning CAUTION Our products are NOT waterproof The equipment...

Страница 16: ...ght 5 4 2 Detaching Parts De construction of the lights should be done following a reverse order of the instructions in section 4 The BH200 is a completely enclosed light and arm it should not be open...

Страница 17: ...connectors between the light and the extension cable are fully connected Poor Light output Power Button Please check the variable intensity is not just set to its minimum To do this hold in the power...

Страница 18: ...SUES phone DARAY S Service Delivery Team on 44 0 800 878 9864 Opening times 9am 5pm Monday to Thursday excluding bank holidays 9am 2 30pm on Friday Or email support daray co uk IF THE LIGHT WAS ORIGIN...

Страница 19: ...ges may be applicable OTHER Any situation which is not covered by any of the above We will always try to help but we cannot normally offer a refund For additional clarification please refer to our ter...

Страница 20: ...Daray s instructions Attachment of fittings and accessories not approved by Daray Repairs modifications carried out by service personnel not approved by Daray Damage caused by chemical corrosion from...

Страница 21: ...Page 20 of 20 Service Notes...

Страница 22: ...Page 21 of 20...

Страница 23: ...Page 22 of 20...

Страница 24: ...DARAY Ltd Edison House Robian Way Swadlincote Derbyshire United Kingdom DE11 9DH 44 0 800 878 9864 44 0 333 321 0973 sales daray co uk www daray co uk...

Отзывы: